F15 Series Fetal & Maternal Monitor Service Manual
Periodic Tests
- 62 -
10.2.4 SpO2 Performance Test
This test checks the performance of SpO
2
measurement.
Required Equipment:
SpO
2
Simulator
1)
Connect the equipment and the monitor as shown below:
2)
Switch on the monitor.
3)
Generate the SpO2 signal 98% and 100% in turn.
4)
Check if the SpO2 value displayed on the screen is accurate. Error within 2% is
acceptable.
10.2.5 NIBP Performance Test
This test checks the performance of NIBP measurement.
Required Equipment:
Blood Pressure Simulator
1)
Connect the equipment and the monitor as shown below:
2)
Switch on the monitor.
3)
Generate the NIBP signal 270/215 (235) and 40/10(20) in turn.
4)
Check if the NIBP value displayed on the screen is accurate. Error within 8 mmHg is
acceptable.
10.2.6 TEMP Performance Test
This test checks the performance of TEMP measurement.
Required Equipment:
Resistance Box
1)
Connect the equipment and the monitor as shown below:
SpO
2
Sensor
SpO
2
Simulator
NIBP Cuff
Blood Pressure
Simulator