– SE-1010 PC ECG, Tango+ Interface Notes 

 

                                                                        Page 4 of 5

 

Q.  The Tango+ monitor returns results of 0/0 after blood pressure measurements.  What 

do I need to do to get a BP reading? 

A. 

There are certain noisy conditions where the Tango+ cannot accurately measure BP.  
When the Tango+ encounters these situations, it returns a reading of 0/0.  Placement of the 
microphone attached to the cuff is critical for reliable operation of the Tango+.  Follow the 
instructions in the 

Cuff Tutorial

 (located on the SunTech Medical website under Products 

 Tango+) for correct microphone placement. Follow steps 1 and 2 in Conducting the 

Stress Test in the User's Guide to provide the best conditions to obtain a measurement. 

 

Q.  Can I use a heart rate or blood pressure simulator to test whether the Tango+ is 

working correctly with my stress system? 

A. 

You cannot use a heart rate or blood pressure simulator to test whether the Tango+ is 
working with your stress system.  The Tango+ monitor requires that the ECG signal and the 
Korotkoff sounds, collected by the microphone in the cuff, originate from the same source, 
meaning the patient. 

 

Q.  I cannot clearly see the Tango+ display.  How do I fix this? 
A. 

If you cannot clearly read Tango+, you can adjust the contrast of the display by following 
these steps: 

1.  When the operating screen is displayed, press the 

SELECT

 button 2 times.   This 

will bring up the 

MAIN MENU

 screen.   

2.  Using the UP or DOWN arrows, highlight 

MONITOR SET UP

 and press the 

SELECT

 button. 

3.  Using the UP or DOWN arrows, highlight 

CONTRAST

 and press the 

SELECT

 

button. 

4.  Using the UP or DOWN arrows, modify the contrast of the screen.  When you are 

finished, and press the 

SELECT

 button to confirm the choice. 

5.  Using the UP or DOWN arrows, select 

EXIT

 twice to return to the operating screen. 

 

Q. 

My Tango+ displays a message, “Please VERIFY CALIBRATION” or “Equipment 
Maintenance and Calibration Required.”  What do I do? 

A.  Verification of Pressure Calibration

 

Equipment Required:

 

1.  Calibrated electronic manometer or equivalent. 
2.  500mL volume or the Orbit-K Adult Plus cuff wrapped around something that will not 

break or crush (no glass). 

3.  Hand Inflation Bulb with bleed valve. 
4.  Tubing, Tee pieces, and miscellaneous connectors or you can order the T-Tube Kit 

(SunTech Part # 98-0030-00). 

Procedure: 

5.  When the operating screen is displayed, press the 

SELECT

 button 2 times.  This will 

bring up the 

MAIN MENU

 screen. 

6.  Using the UP or DOWN arrows, highlight 

MONITOR SET UP

 and press the 

SELECT

 button. 

7.  Using the UP or DOWN arrows, highlight 

VERIFY CALIBRATION

 and press the 

SELECT

 button.  The monitor will close its bleed valves and will display on its 

screen the pressure applied to the patient hose connector. 

Summary of Contents for SE-12 Express

Page 1: ...enables the EDAN Stress System to prompt Tango when it needs a BP measurement and allows the Tango BP reading to be transferred to the EDAN display and reports o Part provided by EDAN o Labeled connection ends connect as followed A Connects to the EDAN Stress System B Connects to the SunTech Tango monitor ECG Trigger Cable provides the ECG trigger signal from the EDAN Stress System to the Tango o ...

Page 2: ...ECT button 2 times This will bring up the MAIN MENU screen b Using the UP or DOWN arrows highlight MONITOR SET UP and press the SELECT button c Using the UP or DOWN arrows highlight STRESS SYSTEM and press the SELECT button d Using the UP or DOWN arrows scroll through the list until SUNTECH is highlighted and press the SELECT button to confirm the choice e Using the UP or DOWN arrows select EXIT t...

Page 3: ...row to move the cursor When the cursor is on General Info on the System Setup interface press to open the General Information Setup interface 6 Set Device Type Device Mode and BP Monitor Note Press Tab or Shift Tab to move the cursor to the Device Type list box and then press the Up or Down arrow to highlight Treadmill or Ergometer Press Tab or Shift Tab to move the cursor to the Device Model list...

Page 4: ... of the display by following these steps 1 When the operating screen is displayed press the SELECT button 2 times This will bring up the MAIN MENU screen 2 Using the UP or DOWN arrows highlight MONITOR SET UP and press the SELECT button 3 Using the UP or DOWN arrows highlight CONTRAST and press the SELECT button 4 Using the UP or DOWN arrows modify the contrast of the screen When you are finished ...

Page 5: ...5 of 5 8 Verify the Tango calibration by manually inflating and checking the manometer against the pressure reading on the Tango display 9 Once the calibration has been completed use the UP or DOWN arrows to select EXIT twice and return to the operating screen ...