Operating manual ВІСТ.418266.006-03 НЕ
24
6.2 Installing bdbg.exe program
To install bdbg.exe, run the Setup.exe file from the installation disk's BDBG-09 folder and,
in response to the setup program’s question, wait for the installation to complete.
All the program files are copied to the directory
<bootdrive>: \ Program Files \ SVC \ bdbg
or to the directory specified during installation.
7 Removing technology software
7.1 To remove (uninstall) the technology software, click
Start
and go to
Control Panel
->
Add or Remove Programs
In the list of installed programs, select
Bdbg
and click
Remove
.
8 Validity check of BDBG.EXE file
8.1 The MD5 hash (checksum) of the BDBG.EXE file version 1.5.0.2 is:
17 15 d4 12 ae 27 d1 b9 39 80 21 22 46 de e8 54
8.2 At any time, the MD5-hash of the BDBG.EXE file can be calculated using the Windows
command line:
certutil - hashfile <bootdrive>: \ Program Files \ SVC \ bdbg.exe MD5
If the calculated MD5 hash does not match with the above one, the BDBG.EXE program has
been corrupted and it needs to be reinstalled.
9 Bdbg.exe program
9.1 After the bdbg.exe program is successfully launched, the program window is displayed
and the
Setup
tab is activated (Figure A.1).
Figure А.1 – Bdbg.exe program window with the active
Setup
tab
Firstly, choose the interface language among English, Russian and Ukrainian.
Further select a COM-port, to which the detecting unit is connected through the adapter.
Select the checkbox on the USB-emulation indicator while operating with the COM-port emulated
via USB. If the wrong COM-port is selected, the corresponding diagnostic messages will be
displayed.
The
Measurement results
group serves to save measurement results on the disk in a text file.
The text file is created in the directory:
<bootdrive>:\Users\<User>\My documents\Bdbg