Econix ET2020 User Manual Download Page 3

 

 

2.

 

Photometer Preparation and Testing 

2.1.

 

Remove the cover from the batteries compartment. Insert 4 AA batteries into the 

compartment, or connect the AC adapter to the power socket. 

2.2.

 

Press on the ON button. On the display will appear “Self –Diagnostic”. The battery status 

and all wavelengths values will be displayed. If the instrument works promptly after 25 sec on 
the display will appear starting wavelength value xxx nm

2.3.

 

If the display says “Defective xxx nm source” (where xxx --- the wavelength of the defective 

source), you have to send the device to the repair service for a replacement. 

2.4.

 

If you will read on the display “Defective photo element” the device should be sent for 

repairing. 

2.5.

 

If “Low battery” will be displayed on the screen it means that there is insufficient voltage. 

In this case you have to replace the batteries. 

2.6.

 

To return to the wavelength selection, press the "CANCEL" button. 

2.7.

 

To turn on the backlight, press and hold the "ENTER" key until the display lights up. Turn  

off the backlight by pressing again and holding down the "ENTER" button until light goes out. 

2.8.

 

To turn the device off, press and hold the "OFF" key for 2 seconds. 

3.

 

Calibration and Measurements 

3.1.

 

Prepare standard solutions for the photometer calibration according to the selected 

measurement procedures. 

3.2.

 

Turn on the photometer by pressing the "ON"

3.3.

 

Once the screen displays "xxx nm" (where xxx --- automatically selected wavelength value of 

the first starting set) find the required wavelength by pressing the «» and «». 

3.4.

 

Transfer 2---3 mL of the background solution to a 10 mm standard cuvette. Insert a cuvette 

with the background solution in the cuvette compartment and put a cap on the cuvette. The 
orientation of the acrylic cuvette transparent planes should be toward the keypad. 

3.5.

 

Press the "ENTER" key. The screen of the instrument will display "Measure background?" 

Press the “ENTER” button again to confirm. The screen will return in a countdown time mode, 
after the screen will display "Background measured." * * 

3.6.

 

Insert the cuvette containing 2---3 mL of a sample solution (first standard) into the cuvette 

compartment. Put a cup on the cuvette.     Press the "ENTER" key. The screen of the instrument 