Ecomatik T-Soil User Manual Download Page 3

T-Soil Sensor                                            

                                                     

 

 

 

3. Put about 30 cm of additional cable in the installation hole, so that the sensor is free of 

any pulling force. 

4. Refill the installation hole with the excavated material. 

5. Strain relieve the sensor cable coming out of the soil as close as possible to the instal-

lation position, e.g. on a peg firmly hammered into the ground. 

You need further assistance?  

In case you need further assistance for installation, please do not hesitate to contact us.  

 

4. Wiring and Logger Configuration 

The T-Soil sensor is compatible with our DL 18 data logger (ordered with stereo plug), as 
well as with a wide range of other available data loggers, e.g. CR1000(ordered with bare 
cable  ends).  However,  note  that  suitable  loggers  have  to  provide  a  precise  and  stable, 
switched (sensor should only be powered 100ms before and during measurements) exci-
tation voltage (V

ex

) of usually 2500 mV.  

 

Connecting the sensor with bare cable ends, e.g. to a CR1000 logger: 

The  sensor  can  be  connected  either  in  differential  (wiring  diagram  a),  or  in  single-ended 
(wiring diagram b) mode. 

 

 

Figure A: Differential wiring 
 
 
 
 

 

Figure B: Single-ended wiring

 

 

Wiring Examples:  

Campbell Data Logger (CR1000) 

This section describes how to connect the T-Soil sensor to the widely used Campbell data 
logger CR1000. If you use another data logger, contact us in case you need further assis-
tance.  The  T-Soil  sensor  can  be  connected  in  differential  voltage  as  well  as  in  single-
ended voltage mode, measurement range must be set to 2500 mV. One CR1000 can rec-
ord  up  to  eight  T-Soil  sensors  in  differential  mode,  or  sixteen  T-Soil  sensors  in  single-
ended mode. 

 

 

 

Reviews: