background image

Application

 

The  ecom-UNO-CO  is  suitable  for  the  measure-
ment of carbon monoxide concentrations in ambi-
ent air.  
 
 
 
 
 
The  operation  range  allows  the  measurement  of 
concentrations  from  0  to  1000  ppm.  Please  al-
ways  observe  the  valid  safety  regulations  (e.g. 
MAK value)! 
 
 
 
 
 

Instrument design 

 

Battery Indicator 

When  the  battery  indicator  is  blinking  (no  point 
anymore:         ), the batteries have to be replaced 
or recharged (rechargeable batteries optional). 
Recharge as follows: 

●Connec

t adapter cable (Option) with car cigarette 

lighter. 

●Connec

t  adapter  cable  (Option)  with  charging 

socket of ecom-UNO-CO. 

●During  the  charging  procedure  an  a

rrow  shows 

on the field „Charge“.

 

●The batteries are recharged in full  when the a

r-

row has disappeared. 
 
 
 
 
 
 

Set Operation Mode 

The ecom-UNO-CO can be operated in 2 different 
modes: 

Standard

 

The  unit  automatically  switches  off  after  ca.  30 
minutes. 

Continue

 

The unit doesn´t switch off automatically. 
The  operation  mode  can  be  checked/selected  as 
follows: 

The  unit  being  switched-off,  press  the  CAL  key 

until  the  display  shows  „S“

  =  Standard  or 

„C“  = 

Continue. 

Press the BEEP key or the COmax key to switch 

the operation mode. 

Press  the  CAL  key  to  perform  measurement  in 

the selected mode. 
 
 
 
 

 

 

 

Switch-On Instrument 

 
 
 
 
 

Press the CAL Key.

 

 
 
 
 
 
 

 

The ecom-UNO-CO starts a calibration phase of 

max. 1 minute (0 blinks on the display). 

After  completion,  the  unit  sets  the  zero  point  at 

the  CO  concentration  in  the  ambient  air  is  ready 
for operation. 
 

Signal Function 

The  unit  warns  by  exceeding  of  a  50  ppm  CO 
concentration with a beep tone. If the CO concen-
tration  increases,  the  repeating  frequency  of  the 
beep  tone  gets  shorter  by  each  signal  threshold 

(see „Technical Data“). By values over 1000 pp

m, 

only dashes are displayed. 
 

Switch-Off Signal Function 

Press the BEEP key to switch off the beep tone. It 
automatically  switches  on  again  after  15  minutes 
or by a CO concentration lower than 20 ppm. 

 

COmax Function 

Press the COmax key to call up the highest value 
detected since the instrument switch-on. Continu-
ous pressing the CO max. key deletes the maxi-
mal value. 
 

 

ecom-UNO-CO  

 

 

 

2

 

ecom-UNO-CO  

 

 

 

3

 

ecom-UNO-CO  

 

 

 

4

 

Hand over defect or used  

batteries to designed  

collection places only!

 

 

Carbon monoxide 

is poisonous and 

can be fatal!

 

 

Display

 

Charging 

socket  

(front side)

 

BEEP 

key

 

 

CAL key

 

Battery com-

partment 

(back side)

 

COmax 

key

 

Battery 

capacity

 

Switch on the instrument by fresh 

air ambient conditions!

 

 

Segmental test is 

performed

 

 

The unit is not suitable 

for personal safety use!

 