FAQ

Bar 14 
(calibration incorrect)

•  A gas concentration was present during  
   zero calibration.
•  Too little difference between zero and  
   1000ppm calibration
•  Sensor defective

Bar 11 
(sensor head) 

•  No sensor head available
•  Sensor heating element defective

Summary of Contents for LSG

Page 1: ...OPERATION MANUAL English Status Nov 2020 Version ecom LSG ...

Page 2: ...ncentration on the display and warns by means of an optical and acoustic signal The sensitivity of the sensor can be adjusted so that even small leaks can be detected The sensor head is connected to a movable flexible tube so that even parts of the installation which are difficult to access can be reached ecom LSG 2x 1 5 V AAA Batteries ...

Page 3: ...NG Hazard of Injury Improper work on gas pipes can result in injuries Work on gas pipes may only be carried out by qualified personnel When working on gas pipes observe the applicable safety regulations CAUTION Damage to the Sensor Substances such as silicones halogenated hydrocarbons lead tetra ethyl sulphur compounds and organic phosphorus compounds lead to partially irreversible sensitivity los...

Page 4: ...6 5 4 3 2 1 Initiation NOTICE NOTICE Hold the On Off key down until you hear a signal 1 Bar no concentration 14 Bars max concentration 1 Switch on the device 2 Wait for the heating up phase 3 3 fold beeps indicate operational readiness 4 Display shows the gas concentration as a bar chart ...

Page 5: ...leak the measured value display reacts NOTICE If the first switching threshold is exceeded the red LED flashes and an acoustic signal sounds With increasing concentration the LED flashes at shorter intervals and the signal tone becomes more intense 3 The highest gas concentration registered so far is represented by a grey bar Deviations of the bargraph display from zero can be set to zero by press...

Page 6: ...ery capacity is displayed NOTICE If 4 bars are shown on the display the battery capacity is sufficient for about 1 hour 3 Use the HOLD key to make the following settings Without lighting without acoustic signal With lighting without acoustic signal Without lighting with acoustic signal With lighting with acoustic signal 4 Press the ON OFF key to save the selected setting and exit the menu Switch o...

Page 7: ...laced Power Supply Battery Change 1 Open the battery compartment on the back 2 Remove old batteries 3 Insert new batteries 4 Close the battery compartment The two 1 5 V AAA batteries must be replaced when a grey battery symbol appears in the top left hand corner of the display The remaining capacity is 1 hour If the voltage drops further the battery symbol turns black and the unit switches off aut...

Page 8: ...itch off After 30 min Operation Temperature 5 C 40 C Storage Temperature 25 C 40 C Operation Moisture 20 80 rF NOTICE Subject to technical changes Status 11 2019 Changing the Sensor 1 Screw on the sensor head 2 Pull off sensor element 3 Install new sensor element 4 Screw on the sensor head NOTICE Switch off the device before replacing the sensor ...

Page 9: ...ion incorrect A gas concentration was present during zero calibration Too little difference between zero and 1000ppm calibration Sensor defective Bar 11 sensor head No sensor head available Sensor heating element defective ...

Page 10: ...ed of in an environmentally friendly manner by an authorised centre NOTICE On request we take back old appliances and dispose of them in an environmentally friendly way To ensure the accuracy of your measuring device we recommend an annu al inspection by an authorised ecom partner In case of heavy use shorter inspection intervals should be chosen please contact your ecom partner ...