15
Range & Accuracy
Range
Min
Max
Accuracy
ohms Read (low)
0.00 Ω
400.00 Ω
± 0.1 Ω
ohms Read (high)
400.0 Ω
1500.0 Ω
± 0.5 Ω
1500.0 Ω
3200.0 Ω
± 1.0 Ω
Range
Min
Max
Excitation Current
Accuracy
Ohms Source (low)
5.00 Ω
400.00 Ω
0.1 to 0.5 mA
0.15 Ω
5.00 Ω
400.0 Ω
0 0.5 to 3 mA
0.1 Ω
Ohms Source (high)
400.0 Ω
1500.0 Ω
0.05 to 0.8 mA
0.5 Ω
1500.0 Ω
3200.0 Ω
0.05 to 0.4 mA
1.0 Ω
Thermocouple Read and Source (All errors included)
Range in °C
Min
Max
Accuracy
J Thermocouple
-200.0
0.0
0.6°C
0.0
1200.0
0.4°C
K Thermocouple
-200.0
0.0
0.8°C
0.0
1370.0
0.5°C
T Thermocouple
-200.0
0.0
0.8°C
0.0
400.0
0.5°C
E Thermocouple
-200.0
0.0
0.5°C
0.0
950.0
0.4°C
R Thermocouple
-20
0.0
2.4°C
0
500
1.7°C
500
1750
1.3°C
S Thermocouple
-20
0
2.4°C
0
500
1.7°C
500
1750
1.4°C
B Thermocouple
600
800
2.1°C
800
1000
1.7°C
1000
1800
1.3°C
L Thermocouple
-200
0.0
0.45°C
0.0
900.0
0.4°C
U Thermocouple
-200
0.0
0.7°C
0.0
400.0
0.45°C
N Thermocouple
-200
0.0
1.1°C
0.0
1300.0
0.6°C
mV Read/Source
-10.00
75.00
0.015 % ± 2 dig.