Eco Sensors
Ozone Analyzer, Model UV-100
3
Division of KWJ Engineering Inc.
User Manual, Rev. 2.0
[email protected] or (800) 472-6626
Front Panel
LCD Display
The data displayed on the LCD depends on a variety of settings that are available via the
Instrument
Menu
and
Function
Knob.
By default, when the instrument starts, the front panel LCD will display the software version number
and the date & time, followed by ozone measurements, temperature & pressure measurements and the
date & time. Also available are average & instantaneous ozone measurements, log sample number and
flow rate.
LED Indicators
> 0.1 PPM
:
Measurement is above the OSHA Limit for 8-hour exposure to ozone
Low Flow
:
Input sample draw is less than the average of 1 LPM. Check the inlet tubing and filter for
obstruction.
Low Lamp
:
UV Lamp output is below normal levels. This can be caused by internal contamination or a
fault in the lamp or lamp circuitry. Contact our Technical Support if this light turns on.
Power
:
UV-100 is powered and operating.
Function Knob
The
Function
Knob is used to navigate the instrument menu. The knob rotates clockwise or
counterclockwise to move the selection cursor around the menu. The knob depresses like a pushbutton
to enter the menu and activate a cursor selection.