OZONE SOURCE 

Model OG-3 

User Manual

 

 

Overview

 

 

The OG-3 Ozone Source may be utilized for calibration verification and approximate field calibration of Eco Sensors instruments.  It 
utilizes  a  precision  UV  micro-lamp  that  emits  0.1  ppm  or  1.0  ppm  of  ozone  (specify  when  ordering).  The  ozone  output  level  is 
marked on the OG-3. The OG-3 is useful for checking instruments in the field to verify their response to ozone. The OG-3 may also 
be  used  for  approximate  field  calibration  of  instruments.  However,  a  field  calibration  with  the  OG-3  does  not  substitute  for  a 
laboratory  calibration  and  is  not  NIST-traceable.  If  a  NIST-traceable  laboratory  calibration  is  desired  or  required  for  your  Eco 
Sensors instrument, please contact us at (800) 472-6626 or e-mail at [email protected].  

 

Operation 

 

Place the OG-3 and ozone sensing instrument on a flat surface in a room free of air currents and extraneous gases and chemical 
vapors, including but not limited to VOCs (volatile organic compounds). 

Connect the AC adapter to the OG-3. The fan will run and the blue light of the UV lamp is visible through the 10 mm hole on the side 
of the instrument. If the fan is not running, ensure the power source is 12 VDC with a minimum current capacity of 300 mA. If the fan 
is running but the lamp is not operating, contact your distributor or Eco Sensors for further assistance. 

Allow  both  the  OG-3  and  ozone  sensing  instrument  to  warm  up  (unconnected  and  separated  by  at  least  12  in)  for  at  least 
30 minutes. 

Insert the sensor of the ozone sensing instrument into the 10 mm hole in the side of the OG-3 (See illustration below). 

The  ozone  sensing  instrument  should  respond  within  1  minute  and  the  reading  will  climb  to  a  steady  value.  Typically,  after  3 
minutes,  the  ozone  sensing  instrument  will  stabilize  and  the  instrument  reading  may  be  noted.  Do  not  leave  the  ozone  sensing 
instrument connected to the OG-3 for more than 4 minutes.  The OG-3 should be disconnected from its power source when not in 
use to prolong the lamp life. 

The UV lamp in the OG-3 should show stable and constant output for  approximately 500 hours of use. Contact your distributer or 
Eco Sensors for a UV lamp replacement and/or recalibration. 

 

 