Appendix A: Calculating the %O

2

 in Air 

 

The oxygen concentration in dry air (zero humidity) is 20.945 and decreases with increasing 

humidity.  The %O

2

 in air can be calculated from the barometric pressure (P

b

, in mmHg), the 

relative humidity (Rh), and the saturated water vapor pressure (P

ws

, in mmHg) by using the 

following formula: 

 

%O

2

 = 20.945% x (P

b

 - P

ws

 x (Rh/100)) / P

b

 

 

The saturated water vapor pressure (P

ws

) is a function of the ambient temperature (T

a

) and is 

given in the table below.  For example, at 21 

C, P

ws

 = 18.65 mmHg. 

 

 

T

a

 

(

C) 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P

ws

 (mm Hg) 

4.579 

4.926 

5.294 

5.685 

6.101 

6.543 

7.013 

7.513 

8.045 

8.609 

10 

9.209 

9.844 

10.518 

11.231 

11.987 

12.788 

13.634 

14.530 

15.477 

16.477 

20 

17.535 

18.650 

19.827 

21.068 

22.377 

23.756 

25.209 

26.739 

28.349 

30.043 

30 

31.824 

33.695 

35.663 

37.729 

39.898 

42.175 

44.563 

47.067 

49.692 

52.442 

40 

55.324 

58.34 

61.50 

64.8 

68.26 

71.88 

75.65 

79.60 

83.71 

88.02 

50 

92.51 

97.2 

102.09 

107.2 

112.51 

118.04 

123.80 

129.82 

136.08 

142.60 

60 

149.38 

156.43 

163.77 

171.38 

179.31 

187.54 

196.09 

204.96 

214.17 

223.73 

70 

233.7 

243.9 

254.6 

265.7 

277.2 

289.1 

301.4 

314.1 

327.3 

341.0 

80 

355.1 

369.7 

384.9 

400.6 

416.8 

433.6 

450.9 

468.7 

487.1 

506.1 

90 

525.76 

546.05 

566.99 

588.60 

610.90 

633.9 

657.62 

682.07 

707.27 

733.24 

 

1 mmHg = 0.01934 lbf/in2 = 1 torr 

 

1 atm = 14.696 lbf/in2 = 760 torr = 101325 N/m2 

 

 

 

Summary of Contents for XCM114

Page 1: ...XCM114 O2 Lambda AFR Module Instruction Manual 07 15 2014 P N XCM114 4...

Page 2: ...this manual may be photocopied or reproduced in any form without prior written consent from ECM ENGINE CONTROL AND MONITORING Information and specifications subject to change without notice XCM114 is...

Page 3: ...nment where an oxidizer and fuel exist because the sensor may ignite the mixture Important Operation Notes 1 Before installing the O2 sensor apply a small amount of non lead containing antiseize compo...

Page 4: ...4 XCM114 Kit Contents Not pictured This instruction manual P N XCM114 4...

Page 5: ...rom 12 to 48 from the exhaust valve s of the engine Mount the O2 sensor in a location where liquids ex condensed water will not collect on the sensor Lightly coating its threads with a non lead contai...

Page 6: ...aged sensors 1 Hold the sensor in quiescent air for at least 20 minutes Measure Vout or Iout and determine the O2 in the air from Appendix A Denote this O2 as O2cal and the measured Vout or Iout as Vc...

Page 7: ...le should be interpreted as 0 i e there are no O2 molecules in the mixture Negative values are given for use in calculating Lambda AFR using the table below where it indicates a rich mixture O2 vs Lam...

Page 8: ...rrected Lambdameasured B x P 1 B x P Equation 3b where Lambdacorrected the Lambda corrected for exhaust pressure Lambdameasured the Lambda output by the AFM B 0 000627 for Lambda 1 0 rich B 0 000830 f...

Page 9: ...nected internally Current output mode using internal Int power i e Current Transmitter Type 3 Current output mode using external Ext power i e Current Transmitter Type 4 1 RDAQ in 4 20mA systems is ty...

Page 10: ...on each side of the module and slide the PCB out of the enclosure 3 The switches for mode and current output power source are located on the lower left of the board labeled S3 S4 Set the position of t...

Page 11: ...ed to the left Do not change the blue potentiometer adjacent to it Type Switches On Vout Range Iout Range 1 2 6 0 5 to 2 5 V 5 to 9 mA 2 2 7 2 5 to 4 5 V 9 to 12 mA 3 2 8 3 5 to 5 5 V 11 to 15 mA 4 4...

Page 12: ...r the integrated diagnostic system Failure Mode Judgment Condition Action After Failure Condition Heater Stop Code Heater Open Off 1 x1 Heater Shorted Off 1 x2 Under or over Vbat Vbat 11V or Vbat 28V...

Page 13: ...press implied or arising by operation of law trade usage or course of dealing are hereby disclaimed The warranty is void if the control module is opened or the wiring harness is modified LIMITATION OF...

Page 14: ...926 5 294 5 685 6 101 6 543 7 013 7 513 8 045 8 609 10 9 209 9 844 10 518 11 231 11 987 12 788 13 634 14 530 15 477 16 477 20 17 535 18 650 19 827 21 068 22 377 23 756 25 209 26 739 28 349 30 043 30 3...

Page 15: ......

Page 16: ...Los Altos CA 94023 0040 USA 408 734 3433 Fax 408 734 3432 www ecm co com...

Reviews: