Performing an Exam
54
KOSMOS User Guide
1.
Tap the More options icon next to the color map on the right side of the
screen.
2.
Select the color map you like.
3.
To invert the color map, select the check box, and tap
OK
to save the
changes.
Image mode controls
Flipping an image
You can only flip an image when you are scanning the heart.
To flip the image, double-tap the orientation marker.
Adjusting body mass index (BMI)
In KOSMOS, BMI is used to adjust the penetration level. BMI is not calculated or
supplied from the patient's height or weight entry.
There are three levels of adjustment:
•
Low (default)
•
Medium
•
High
When you adjust the BMI, it changes the penetration signal for the ultrasound
parameters, so if you have a patient with a high BMI, you will want to increase
the BMI (and vice versa).
To adjust BMI, tap
BMI
, and select one of the three different penetration
levels.
Adjusting depth and gain
To adjust depth:
•
To increase the displayed depth, turn the Depth knob clockwise.
•
To decrease the displayed depth, turn the Depth knob counter-clockwise.
Summary of Contents for KOSMOS
Page 1: ......
Page 2: ...This page left intentionally blank ...
Page 18: ...KOSMOS User Guide 10 KOSMOS capabilities This page left intentionally blank ...
Page 42: ...Using KOSMOS 34 KOSMOS User Guide This page left intentionally blank ...
Page 75: ...Reviewing an Exam 68 KOSMOS User Guide This page left intentionally blank ...
Page 89: ...KOSMOS Maintenance 82 KOSMOS User Guide This page left intentionally blank ...
Page 117: ...Safety 110 KOSMOS User Guide This page left intentionally blank ...
Page 130: ...KOSMOS User Guide 123 This page left intentionally blank ...