28
DISABLE
turns off the relay and removes it’s icon from the HOME screen
Setting up an Alarm Relay
Choose
ALARM
Press
CHANGE
to enter new values
Choose ON Set Point, Press OK
Enter value using
▲
or
▼
and NEXT to move to the next digit, press OK, press BACK (Min –Max values
indicate the range of acceptable values)
Choose Expiration, Press OK,
Choose time from drop down menu using
▲
or
▼
, press OK, press BACK
Choose OFF Set Point, Press OK
Enter value using
▲
or
▼
and NEXT to move to the next digit, press OK, press BACK
Choose Delay ON, Press OK
Enter value using
▲
or
▼
and NEXT to move to the next digit, press OK, press
BACK
Choose Delay OFF, Press OK
Enter value using
▲
or
▼
and NEXT to move to the next digit, press OK, and
press BACK when done to exit Relay 1.
Repeat for Relay 2 and Relay 3.
3.3.4
E
XIT
M
ENUS AND
R
ETURN TO
M
AIN
D
ISPLAY
Press HOME Key to return to the Home Menu Screen
Press Hold to turn OFF Hold
Press EXIT Key to exit the menu
“Save Changes?” press YES
Choose Display Mode,
DATA
,
mV
or
GRAF
by pressing selection Key. The
selection key displays which screen will be displayed next.
o
The type of graphical display used, Line, Bar or Gauge is selected in
CONFIG → XMTR → LCD → GRAPH → LINE, GAUGE, BAR
3.3.5
S
ENSOR
S
TART
U
P
All sensors are supplied with protective caps over the sensing end. Remove the cap(s) from the sensor before
installing in the process. All sensors were calibrated at the factory before shipment, no calibration should be
necessary before use. Allow the sensor to equilibrate to the process solution conditions for ½ hour before
verifying the reading against a grab sample. If calibration is required follow the instruction in Section 4.0 below.
3.4
USER
SELECTABLE
OPTIONS
3.4.1
S
CREEN
L
IGHTING
LED back lighting is available on AC and DC powered instruments only.
Contrast can be adjusted for optimal viewing. The Backlight can be adjusted to timeout after a set period of time
or remain on.
Location:
CONFIG → XMTR → LCD → Set Up → CONT, BACK LIGHT