ECC X-RAY 870 Instruction Manual Download Page 5

 

Exposure Time Measurements:

 

The exposure time will be displayed in milliseconds (ms). 

 

 
 

mA Lead Connection

 

 

The mAs meter was designed to be connected in the ground return line of the high 
voltage transformer and must be in the actual current path of the x-ray tube. The 
meter can be connected in series with the rectified tube current (DC Path), via the 
DC input jacks or it can be connected using the AC input jack in series with the 
transformer downstream of the rectifier (AC Path). However, all tube current must go 
through the mAs Meter to have an accurate reading. If an alternate or parallel 
current pathway of the x-ray generator is present, it could bypass the meter and 
invalidate the reading. Do not use resistors, varistors, bypass, or distributed 
capacitors in parallel with the meter. This will affect calibration. Before using the mAs 
meter, examine the circuit of the x-ray generator and be certain that the mAs meter 
is connected so that such components will not affect the readings. Varistors, which 
have limiting voltage above 10V, will probably not affect the reading. 

 

STATUS MESSAGES

 

Applies to kVp and mA Measurements

 

 

The alphanumeric display on the Model 870 is capable of displaying messages 
along with the output readings that give the user additional information.  Each 
message is described below: 

 

Rdy 2A- Unit is ready for a reading in the 2 Amp range. 

Rdy 200mA- Unit is ready for a reading in 200mA Range 

Over Range - Over Range.  The x-ray output is too high.  Switch to 2 Amp range. 

 

Low Battery - Battery is low.  The unit may still be operated for several hours, but 
accuracy may suffer.  Replace the batteries as soon as possible. 

 

 

BATTERY REPLACEMENT

 

 

The Model 870 is supplied with one 9V battery.  Replace the battery when the "Low 
Battery" indication shows in the display.  Any standard 9V batteries can be used in 
the 870.  We recommend Duracell MN1604 9 Volt Alkaline Battery  To prolong 
battery life, turn the instrument off when not in use. 

 

To replace the battery, open the battery compartment on the bottom of the case 
using a small screwdriver or thumbnail.  Remove the old batteries, and install the 
new batteries.  Observe the correct polarity of the battery as shown on the bottom of 
the battery compartment. 

Summary of Contents for 870

Page 1: ...mAs Meter Rev June 24 2013 Electronic Control Concepts PO Box 182 1160 US Route 50 Milford Ohio 45150 9705 USA Phone 800 VIP XRAY Fax 845 247 9028 sales eccxray com www eccxray com A Division of US Nu...

Page 2: ...mAs Meter INSTRUCTION MANUAL TABLE OF CONTENTS Page DESCRIPTION 3 GENERAL INSTRUCTIONS SWITCH SETTINGS 4 mA mAs OPERATION 4 STATUS MESSAGES 5 BATTERY REPLACEMENT 5 WARRANTY 6 SERVICE INFORMATION 6 SP...

Page 3: ...y for the user to program correction factors for different types of x ray waveforms The Model 870 automatically resets at the beginning of each exposure holding the reading until the next exposure The...

Page 4: ...ug mA Cable While Unit is ON Line up the pins in the cable and chassis connecter before inserting When taking mA or mAs measurements there is no minimum exposure time requirement For AC x rays we sugg...

Page 5: ...have limiting voltage above 10V will probably not affect the reading STATUS MESSAGES Applies to kVp and mA Measurements The alphanumeric display on the Model 870 is capable of displaying messages alo...

Page 6: ...ing the first year after shipment that does not show obvious signs of abuse Contact the factory as described below SERVICE INFORMATION If a unit should need calibration or service please contact the f...

Page 7: ...e POWER 9 Volt battery accessible from bottom of case Low battery indicator BATTERY LIFE 100 hours continuous Typically over one year of normal use WARRANTY 2 years from ship date PHYSICAL SIZE 80 X 1...

Reviews: