
ebro Electronic GmbH
Peringerstrasse 10
85055 Ingolstadt, Germany - www.ebro.com
Ref.No.
:
082011
Ref.No.
:
082011
Ref.No.
:
082011
TFI-54
Thermometer Operating Instructions
The thermometer is a non-contact infrared thermometer. There are many mathematical modes for the Infrared function. Please remember to
keep away from children and don‟t use it for safety related applications.
(Default Screen)
Simply aim the thermometer at the
measure target with Lens (
○
,
9
) and
press Meas. key (
○
,
7
) to display the
surface temperature. The
Distance:Spot is 12:1. Please make
sure the target area is within the
field of view.
FUNCTION
Press Mode key (
○
,
2
) for scrolling more display function as follows.
Here will show the emissivity data. (The default emissivity is 0.95.)
Press Mode key (
○
,
2
) , then press Lock key (
○
,
4
) or
o
C/
o
F key (
○
,
3
) to
set the emissivity
, then press Mode key (
○
,
2
) to
confirm it. The emissivity can be changed from 0.10 (10E) to 1 (100E).
Press Mode key (
○
,
2
) for the
Maximum (MAX), Minimum (MIN), Different between MAX and MIN (DIF) and Average (AVG)
modes. During the measurement, the special modes reading will be displayed beside the mode icon.
Press Lock key (
○
,
4
)or
o
C/
o
F key (
○
,
3
) to change the
High Alarm (HAL)
or
Lo Alarm (LAL)
, then press Meas. key (
○
,
7
) to
confirm it. When the reading is outside the
High Alarm (HAL)
or
Lo Alarm (LAL)
limit. The High or Low icon will flash and you
will hear a beep sound.
ADD VALUE
In E,MAX, MIN, DIF, AVG
mode:
Press Lock key (
○
,
4
) for
LOCK mode
ON/OFF.
The lock mode is particularly useful for continuous monitoring
of temperatures for up to 60 minutes.
Press
o
C/
o
F key (
○
,
3
)
for
o
C
or
o
F transferred.
In all modes: First hold on
the Meas. key (
○
,
7
)
and press
Lock key (
○
,
4
) for
backlight function
ON/OFF.
and press
o
C/
o
F key (
○
,
3
) for
laser function
ON/OFF.
Memory
Recall
the last reading memory
by pressing Mode key (
○
,
2
) while power off.
CAUTION
1.
WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT.
2.
USE EXTREME CAUTION WHEN OPERATING THE LASER.
3.
NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4.
KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times,
care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol, allowing the lens to fully dry
before using the thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
„Hi‟ or ‟Lo‟ is displayed when the temperature being measured is outside of the settings of HAL and LAL.
「
」
「
」
„Er2‟ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. „Er3‟ is displayed
when the ambient temperature exceeds 0
o
C (32°F) or +50
o
C (122°F). The thermometer should be allowed plenty of
time (minimum 30 minutes) to stabilize to the working/room temperature.
「
」
Error 5~9
, for all other error messages it is necessary to reset the thermometer. To reset it, turn the instrument off,
remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message remains
please contact the Service Department for further assistance.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
„Battery OK‟: measurements are possible
„Battery Low‟: battery needs to be
replaced, measurements are still possible
„Battery Exhausted‟: measurements
are not possible
BATTERY REPLACEMENT
1. Pull out the rubber gasket (
○
,
10
)
2. Loosen the screw and pull out the battery cover
3. Replace and reinstall with new batteries
4. Place the battery cover back and fasten the screw
5. Plug the rubber gasket (
○
,
10
) to avoid failure of water proof
When the „Low Battery‟ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Measurement Range
-60 to 550
o
C (-76 to +1022
o
F)
Operating Range
0~50°C (32~122°F)
Accuracy, (Tobj=15-35°C,Tamb=25°C)
+/-1.5°C (2.7°F)
* IP54 Splash Proof