background image

BELT INSPECTION AND REPLACEMENT: 

 
Inspect  belt  periodically  and  replace  if  broken  or  if  the  cords  are  exposed  or 
frayed.  Small outer fabric breaks are not harmful. 
 
Removal of the drive belt may require the adjustment or removal of the motor 
and/or the large drive pulley. Installation of a new belt is as follows: 
 

1.  Unplug unit to avoid shock and/or electrical component damage. 
2.  Remove the left side panel to expose the motor and other components.  
3.  Take note or mark the location of the motor along the adjustable slots in 

base plate.  

4.  Loosen  the 

socket  (“allen”)  screws  that  fasten  the  motor  to  the  base 

plate. 

5.  Slide the motor towards the large pulley to allow the belt to slacken. 
6.  Remove belt and install new belt with identical number of teeth. 
7.  Adjust tension on belt. 

Note:

 cogged style belt does not need excessive 

tension.    

    

 
 

MOTOR: 

 
In the case of a high duty cycle, brushes should be inspected and/or replaced 
at determined intervals (12-24 months).  Brushes can be removed by carefully 
following the steps outlined below:  

1.  Remove  the  left  side  panel.  Once  disassembled  you  can  perform  the 

maintenance to the brushes.  

2.  Unscrew the black plastic brush holders located on each side of the motor.  

Be careful not to lose the inside spring when unscrewing brush holders. 

3.  Inspect  brushes  and  clean  or  replace  as  necessary.    Replace  brushes  if 

worn to less than 1/8

”.  

 

If you have any doubts or inquiries concerning operation contact your supplier or 

Eberbach Corporation

 technical service. 

Summary of Contents for EL680.Q

Page 1: ...Schooner Street Belleville MI 48111 USA TEL 800 422 2558 FAX 734 665 9099 www Eberbachlabtools com EL680 Q QUENCHERS Hand Motion Shaker Variable Speed with Timer and Tachometer 100 600 2 opm 115 V 50...

Page 2: ...ON THE UNIT SHOULD BE PERFORMED BY QUALIFIED TRAINED PERSONELL ONLY INITIAL PREPARATION Place shaker on a level table or bench NOTE This unit should be operated by trained personnel With the power swi...

Page 3: ...equipment without first reading this manual 3 This manual should always be attached to the equipment and made available to all users 4 If you have any doubts or inquiries please contact your supplier...

Page 4: ...kg 8 8 lbs Oscillations per minute 100 600 2 Power Consumption 0 7 Amps INSTALLATION 1 Place the apparatus on a flat horizontal level surface with enough stability to resist its own vibrations 2 Fit...

Page 5: ......

Page 6: ...ad 4 Enter Button 5 Cursor Operation Press the Power Switch down into the ON position The LCD should light up and show the main menu There are three different run modes to choose from Select a run mod...

Page 7: ...l increase decrease Once engaged the timer will begin counting all the way up to 999 hours before the timer overflows back to zero To pause the shaker move the Cursor to the PAUSE item by pressing the...

Page 8: ...s on the D PAD can be used to move the cursor between hours minutes or seconds and the up and down arrows can be used to set the time The longest run time allowed is 999 59 59 Once the time is set mov...

Page 9: ...shown in parenthesis is the target speed controlled by the user and the speed listed to the right is the actual speed as read by the tachometer The shaker and countdown timer can be paused and resumed...

Page 10: ...e up down arrows on the D PAD The left right arrows can be used to move the Cursor between START and EXIT EXIT will take the user back to the main menu Keep the Cursor next to the START line item and...

Page 11: ...press and hold the green Enter Button Next the user must turn the Power Switch on while continuing to hold the Enter Button After a few seconds the user will be presented with the utilities menu Navi...

Page 12: ...e model numbers Once the user has selected the correct model number press the right arrow button on the D PAD Use the up arrow button on the D PAD to change the motor setting to Bodine Navigate the Cu...

Page 13: ...controlled by the user The shaker will attempt to match this speed and should do so within ten seconds after making an adjustment The rightmost speed is the tachometer reading which shows the user wh...

Page 14: ...use warm soap and H2O solution FUSE REPLACEMENT The motor fuse is a 1 25A 1 25 x0 25 Time Delay Fuse The line fuses are 3A 5x20mm Time Delay Fuse Preventative maintenance has set the fuse to blow if...

Page 15: ...arge pulley to allow the belt to slacken 6 Remove belt and install new belt with identical number of teeth 7 Adjust tension on belt Note cogged style belt does not need excessive tension MOTOR In the...

Page 16: ...25AMP TIME DELAY 1 6113 FUSE TIME DELAY 3A 2 6131 1 FUSE HOLDER 1 6217 FEMALE QUICK DISCONNECT 1 6220 CABLE TIE HOLDER WITH ADHESIVE BACKED 3 6267 2 CONNECTOR 1 6278 AC INLET 1 6333 MOTOR CONTROLLER 1...

Page 17: ...ADE 5 HEX NUT 2 9435 12 SAE WASHER 2 9950 BLUE LOCTITE 242 1 EL680 Q 009 MOTOR CONTROLLER SPLASHGUARD 1 ELP680 HD 211 RIGHT SIDE COLUMN 1 ELP680 HD 212 PULLEY BLOCK 1 ELP680 HD 217 LEFT SIDE COLUMNS 1...

Page 18: ...1 ELP680 SS 220 BASE PLATE 1 ELP680 SS 223 SIDE PANEL 2 ELP680 SS 508A CASE ASSEMBLY 1 If you have any questions or inquiries concerning installation or ordering of replacement parts please contact yo...

Page 19: ......

Page 20: ......

Reviews: