BELT INSPECTION AND REPLACEMENT:
Inspect belt periodically and replace if broken or if the cords are exposed or
frayed. Small outer fabric breaks are not harmful.
Removal of the drive belt may require the adjustment or removal of the motor
and/or the large drive pulley. Installation of a new belt is as follows:
1. Unplug unit to avoid shock and/or electrical component damage.
2. Remove the left side panel to expose the motor and other components.
3. Take note or mark the location of the motor along the adjustable slots in
base plate.
4. Loosen the
socket (“allen”) screws that fasten the motor to the base
plate.
5. Slide the motor towards the large pulley to allow the belt to slacken.
6. Remove belt and install new belt with identical number of teeth.
7. Adjust tension on belt.
Note:
cogged style belt does not need excessive
tension.
MOTOR:
In the case of a high duty cycle, brushes should be inspected and/or replaced
at determined intervals (12-24 months). Brushes can be removed by carefully
following the steps outlined below:
1. Remove the left side panel. Once disassembled you can perform the
maintenance to the brushes.
2. Unscrew the black plastic brush holders located on each side of the motor.
Be careful not to lose the inside spring when unscrewing brush holders.
3. Inspect brushes and clean or replace as necessary. Replace brushes if
worn to less than 1/8
”.
If you have any doubts or inquiries concerning operation contact your supplier or
Eberbach Corporation
technical service.