25
Problem
Solution
Error Message “86” and Ebb Shuts Down
Fluid cartridge is out of fluid
Check the icons on the display to see if the water droplet icon and replace words
are illuminated.
Fluid cartridge is not installed correctly
Ensure the fluid cartridge is installed in Ebb correctly and the cartridge knob is
locked in place.
Error Message “87” and
Ebb
Shuts Down
Ebb has been left in hot environment
Allow Bedside Controller to cool down before re-use. Disconnect power and then
reconnect to restart Bedside Controller.
Fan inlet is blocked
Remove objects and allow Bedside Controller to cool before re-use. Disconnect
power and then reconnect to restart Bedside Controller.
Air outlets are blocked
Remove objects and allow Bedside Controller to cool before re-use. Disconnect
power and then reconnect to restart Bedside Controller.
There is a problem with the Bedside Controller
Disconnect the power cord and then plug back in. If error code repeats, contact Ebb
HumanCare for additional support.
Error Message “88” or “82” and
Ebb
Shuts Down
There is a problem with the Bedside Controller
Disconnect the power cord and then plug back in. If error code repeats, contact Ebb
HumanCare to arrange for additional support.