Icon
Operation
status
Possible cause
Action
In AVR
mode
The UPS is operat-
ing normally but
the utility voltage
is outside normal
mode thresholds.
The UPS is power-
ing the equipment
through an automatic
voltage regulation
device. The equip-
ment is still protected.
On Battery
1 beep
every 10
seconds
A utility failure
has occurred and
the UPS is in bat-
tery mode.
The UPS is powering
the equipment with
battery power. Pre-
pare your equipment
for shutdown.
End of
backup
time
1 beep
every 3
seconds
The UPS is in bat-
tery mode and
the battery is run-
ning low.
Shut down the equip-
ment.
Description
3
SOMATOM | Version 1
17