Eaton Crouse-Hinds MTL Z1920C Instruction Manual Download Page 11

INM MTL 130-0175 Rev 4

7

Slowly open the flow valve beneath the flow meter until a flow of between 150 and 250 ml/min 
is flowing. 

WARNING

To avoid serious and dangerous damage to the catalyst and sensor the sample 

flow must not exceed 250ml/min (0.5cuft/hr).

The instrument will respond within a few seconds to show the oxygen concentration.  

NOTE:

 Combustion of the sample generates heat and it is possible, when the sample is first 

applied, that the catalyst temperature may go higher than its normal control point before 
the controller is able to react.  This will cause the controller to show error condition and the 
relay contacts to go a fault condition.  Whether this occurs or not will depend on the sample 
composition and its flow rate.  The results during this transient state are valid.  It is suggested 
that any external equipment (PLC’s etc) used to monitor the analyser are programmed so as to 
ignore this situation for approximately 60 seconds after the sample has been applied.

 Before switching the analyser off following its use on combustible samples ALWAYS pass air 
through the analyser for a few minutes to purge out the moisture etc. formed when combusting 
the sample.  

4.5  Programming

The user programmable features are accessed by pressing and holding the Edit button for 
approximately 8 seconds when the instrument is in normal measurement mode.   
Each subsequent momentary press of the Edit button scrolls the display through a sequence of 
adjustable parameters.  The order of the sequence is as follows:

Alarm 1 Set point   Alarm 1 Mode   Alarm 1 Hysteresis   Alarm 2 Set point   Alarm 2 Mode 

 Alarm 2 Hysteresis   Analogue Output top scale value   Return to measurement mode.

NOTE

Only the LED indicators are available for the concentration alarms.  No relay output is 

provided on this model.

 

ALARM 1 - Set point

This is the first parameter to be displayed when the Edit button is pressed and held for eight 
seconds. The display will show 

“E xxxx”,

 and the Alarm 1 LED will flash. 

“xxxx”

 is the value 

of the alarm set point and the flashing LED indicates that it is the set point of alarm 1.  To the 
right of the display a % or ppm symbol is displayed to indicate the units of the value displayed.  
To change the value of the set point, press the up/down arrows until the display shows the 
required value. Observe the 

“%”

 and 

“ppm”

 symbols to ensure the correct setting.   When 

the buttons are first pressed the value changes slowly and the least significant number can 
be changed by giving the buttons a short

 “click”.

  If the button is kept depressed the rate 

accelerates rapidly to enable large changes in value to be achieved quickly.   Once the required 
value is displayed, press the Edit button once to store the value in the analyser’s memory. 
Simultaneously the display will change to show 

“E    x”

 and the Alarm 1 LED will continue to 

flash.

Summary of Contents for Crouse-Hinds MTL Z1920C

Page 1: ...Z1920C MTL Oxygen combustion analyser February 2017 INM MTL 130 0175 Rev 4 Instruction manual MTL gas analysers systems...

Page 2: ...LARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods However you can find a copy of the latest version at http w...

Page 3: ...ectronics 3 3 2 2 Catalytic Reactor and Sensor unit 3 3 3 Electronic unit connections 5 4 OPERATION 6 4 1 First Checks 6 4 2 Start up 6 4 3 Calibration check 6 4 4 Commisioning 6 4 5 Programming 7 5 M...

Page 4: ...iv INM MTL 130 0175 Rev 4 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 5: ...es a signal proportional to the oxygen content of the gas which is processed by the microprocessor control unit The control unit also provides a digital display and a programmable analogue output typi...

Page 6: ...ax Electronics 50 C Max 2 3 Display Multi digit LCD character height 12 7mm 2 4 Display ranges Display range resolution 0 1 from 10 0 to 100 0 01 from 0 01 to 9 99 2 5 Analogue outputs isolated 0 to 5...

Page 7: ...l details and observe the followings points WARNING Parts of the assembly behind the heat shield run at elevated temperatures Care must be taken to ensure that they cannot be easily touched in normal...

Page 8: ...15 16 C a ta ly s t Te mp T C 20 v A C 20 v A C O x yge n S ens or T emp T C O x yge n S ens or S igna l O x yge n S ens or H ea te r P owe r C a ta ly s t He ater P ower Not Us ed Not U s ed S ens or...

Page 9: ...activate a sample cut off in the event that excessive sample flow greater than 250ml min causes the catalyst to over heat Typically the contacts would be used as a logic input to the overall burner c...

Page 10: ...n no combustible gas The oxygen in nitrogen mixture may be obtained in cylinders from a number of gas suppliers The cylinder will need to be fitted with a regulator The calibration gases are introduce...

Page 11: ...when the instrument is in normal measurement mode Each subsequent momentary press of the Edit button scrolls the display through a sequence of adjustable parameters The order of the sequence is as fo...

Page 12: ...ns to set the required value and then Edit button to store the value The display will then change to repeat the above sequence for alarm 2 indicated by the alarm 2 LED flashing ANALOGUE OUTPUT Top sca...

Page 13: ...mmissioning it is advisable to check the calibration as described in section 4 3 after one week of continuous operation Following this check monthly checks should be adequate When carrying out the cal...

Page 14: ...d by a number The meaning of these messages is as follows HELP 1 After finding user calibration data corrupt the backup was found to be corrupt too Return to factory HELP 2 Will not be seen Factory ca...

Page 15: ...nd 13 This should be 4 10 If either test produces abnormal readings then the catalytic reactor assembly will need to be replaced proceed to section 6 5 for instructions on how to remove the catalytic...

Page 16: ...rozavodskaya Str 33 Building 4 Moscow 107076 Russia Tel 7 495 981 3770 Fax 7 495 981 3771 E mail mtlrussia eaton com SINGAPORE Cooper Crouse Hinds Pte Ltd No 2 Serangoon North Avenue 5 06 01 Fu Yu Bui...

Reviews: