![Easy@Home EMT-021 User Manual Download Page 2](http://html1.mh-extra.com/html/easyathome/emt-021/emt-021_user-manual_3592018002.webp)
symbols in sequence: 188.8°F/°C, the last measurement (if the last temperature
is lower than 89.6°F /32.0°C, it will be a Lo°F/Lo°C), the default setting
temperature 98.6°F or 37 °C, then the current temperature of the flexible tip
with a flickering °F/°C (if the ambient temperature is lower than 89.6°F /32.0°C,
it will be a Lo with a flickering °F/°C) to indicate you are ready to test.
NOTE
:
The Lo with a flickering °F/°C is not an indication of low battery.
A true low battery symbol is an empty flickering battery. When the
environment or ambient temperature is higher than 109.2°F(42.9°C), LCD
will display Hi °F(Hi °C), in this case, you may need to wait till the temperature
is lower than 109.2°F(42.9°C) to take.
Use alcohol to disinfect the probe then choose the method you prefer to
measure:
3.
Oral use: Place the probe under tongue and close the mouth for approximately
25 seconds to finish a measurement. Normal temperature range for an
adult is 97.3 º F~98.9 º F (36.3°C ~37.2°C), as it applies to an oral read.
3.1.
when “ ” symbol flash in the corner of the display screen.
Axillary use: Wipe armpit with dry towel then place the probe in armpit firmly
for approximately 25~37 seconds to finish a measurement. Normal temperature
3.2.
Use a non-metal pointed tool to remove the battery
and insert the new battery with the positive (+) side facing up.
Carefully slide the battery cover back into place.
Notes:
Rectal use: Lubricate the probe with water-soluble jelly such as K-Y
jelly. Do not use petroleum jelly like Vaseline for easier insertion because this
will adversely affect the accuracy of the result. Insert the probe into rectum
about 1/2 inch at most (
ensure not too deep to avoid injury
) for about 25
seconds to complete the measurement. Normal temperature range for an
adult is 97.7 ºF ~99.8 ºF (36.5°C ~37.7°C), as it applies to rectal read.
A "Bi Bi Bi" alarm will sound a total of ten times indicating the completion of the
temperature in the normal range.
The power will be shut off automatically if there is no temperature rising for 10
minutes.
Replace the battery
:
Suitable for EMT-021
The battery in your Easy@Home digital thermometer needs to be replaced
To replace the battery
Push or slide the battery cover away from the unit.
3.3.
4.
4.1.
5.
5.1.
5.1.1.
5.1.2.
5.1.3.
range for an adult is 96.8 ˚F ~98.6 ˚F (36°C~ 37°C), as it applies to an
auxiliary read.
Reading Display
Celsius Display
1. It is suggested to stay with the same measurement method. This means
especially not take oral or armpit measurements after taking rectum
temperatures for hygienic reasons.
2. Press the power button for approximately 10 seconds until the unit flickers
in the shutdown state. Then, release the power key and then quickly press
the power key again for a short time to switch between Fahrenheit and
Celsius. Finally, turn the power button off and then back on, and the switch
is successful.
Low Battery Indicator
Fahrenheit Display
Definition of Display Symbols
6.
This thermometer is guaranteed for 1 year from the purchasing date under
normal use. The warranty does not cover the damage of improper use or the
battery running out.
Note:
Any solution, dilution or methods other than mentioned above might
cause reliability problems with the thermometer.
Users should measure, change batteries under normal circumstances, and maintain
the device and its accessories according to the user manual.
WARNINGS
LIMITED WARRANTY
Do not perform maintenance while the thermometer is in use.
The device is not for use in an OXYGEN-RICH ENVIRONMENT.
Before each use, check the device. Do not use a damaged device or device with
damaged electrical wiring. The continuous use of a damaged unit may cause
inaccurate results or bodily injury.
If you have any inquires or concerns about the device, such as setting up,
maintaining or using it, please contact our Customer Service Team.
Do not open or repair the device by yourself.
The main material of the case is ABS. Be cautious of potential allergic reactions to
this materials.
The typical service life of an unused new battery is 100 hours continuous
operation.
Protection against electric shock: built-in powered ME equipment.
Protection against harmful ingress of water or particulates: IP22.
1.
2.
3.
4.
5.
6.
7.
8.
9.
Qu
es�o
ns
o
r c
o
mments?
please call toll-free:
1-855-822-6999 M-F 9 a.m.-5 p.m. CST
E-mail:[email protected]
Easy Healthcare Corpora�on
360 Shore Dr.Burr Ridge,IL USA 60527
PRECAUTIONS
Do not bite, bend or dismantle the thermometer.
Keep the device out of the reach of children/pets to avoid inhalation or swallowing
of small parts.
Do not expose the device to direct sunlight, high temperature or moisture.
Do not sterilize the device. Prevent saliva or cleaning solution from penetrating the
display window.
Clean the thermometer before and after use with soft cloth and a cleaning solution
and store in the plastic case. It is recommended to wipe with an alcohol pad for
home use.
If the LCD display shows the low battery symbol or is not a clear visual as usual,
it means the battery may be running out. Please be sure to replace the battery
before next use in this case.
2.
3.
4.
5.
6.