4.3.1    Measurement of Gain 
 
1.     Equipment Required (or equivalent): 
 

a) 

Osilloscope - Tektronix T921 

b) 

Sweep/Generator -  HP8601A 

c) 

Signal Generator -  Exact Model 7060 

d) 

50  ohm  Detector -   Wavetek  D151 

e) 

Attenuator,  30 dB,  500  Watts  Bird 

 
 

 

 
Figure 4-1.  Gain Measurement 
 
2. Connect the equipment as shown in Figure 4-1, then proceed as follows: 
 

a) 

Set  the  oscilloscope  to  DC,  Time/cm  to  Ext.  X,  and  gain  to 
l0mv/cm. 

b) 

Set  the  sweep  generator  to  the  video  sweep  mode  with  the  start 
frequency at 300 kHz and the sweep width to 35 MHz. 

c) 

Disconnect  the  A150  from  the  set-up  and  connect  the 
sweep/generator RF output directly to the 30 dB attenuator. 

d) 

Adjust  the  output  level  of  the  sweep/generator  for  full  vertical 
deflection on the oscilloscope face. 

e) 

Calibrate  the  scope  face  to  show  3  dB  in  1  dB  steps  by 
attenuating the sweep/generator in 1 dB. 

Summary of Contents for A150

Page 1: ...hout notice August 2012 Revision E 1 A150 Broadband Power Amplifier HIGH RF VOLTAGES MAY BE PRESENT AT THE OUTPUT OF THIS UNIT All operating personnel should use extreme caution in handling these volt...

Page 2: ...lly operational such that it performs according to the original specifications free of charge to the original purchaser Should warranty service be required the unit must be returned to E I freight cos...

Page 3: ...M FM SSB pulse and any complex modulation signals The amplifier has 55 dB gain it is unconditionally stable and will not oscillate even with combinations of mismatched source and load impedance It is...

Page 4: ...m input signal 1 0 Volts RMS for all output load conditions without damage POWER OUTPUT METER True RMS power detection 3 of full scale accuracy POWER REQUIREMENTS 100 240 VAC 47 63 Hz SIZE 5 3 x 16 5...

Page 5: ...1 V rms indicated maximum input voltage is 5 times the level of the input signal required to achieve maximum output Input voltages in excess of 2 volts peak may permanently damage the instrument iii...

Page 6: ...icted If the fault persists please contact Field Service 2 3 3 RS 232 Interface The A150 features a standard RS 232 serial interface suitable for connection to a PC or host system The communication pr...

Page 7: ...nt end The output signal of the MMIC is coupled to the gate of transistor Q1 The further amplified signal appearing at the drain of Q1 is coupled to the input of Q2 This is transformed to 50 and fed t...

Page 8: ...AC_line 110VAC_neutral earth_gnd SELV 5V Inhibit SELV 5V ret Inhibit_ret DC_ok DC_ok_ret 48V_ret 24V 24V_ret 48V A1 Fan 24V Ret 1 12 05 P1 5 9 4 8 3 7 2 6 1 J2 24V_ret 48V_ret RS232 Control Port J5 AC...

Page 9: ...a power level of 13dBm This chapter therefore deals only with certain fundamental procedures for fault location S1 VHF_splitter_20W RF_in RF_out1 RF_out2 50 ohm 34dBm 48V_ret 48V_ret 24V_ret 2100L SC...

Page 10: ...or c Adjust the output level of the sweep generator so that a 50 ohm video detector connected at the output of the unit will not be damaged by excessive power output Reference section 4 4 1 for set up...

Page 11: ...in Measurement 2 Connect the equipment as shown in Figure 4 1 then proceed as follows a Set the oscilloscope to DC Time cm to Ext X and gain to l0mv cm b Set the sweep generator to the video sweep mod...

Page 12: ...rsus frequency sweep on the oscilloscope The average gain should be greater 50 dB The gain variation should be within the 3 dB as shown on the oscilloscope 4 4 2 Measurement of Harmonics Figure 4 2 1...

Page 13: ...prevent movement Provide additional support for projecting parts to relieve these of unnecessary shock Close the carton securely and seal with durable tape Mark the shipping container FRAGILE to ensur...