Menu→Measurements→Draft

 

OK 

Menu→Measurements→CO air

 

AA CTA03A 

E500 

S.n.:0 

Fw:1.00 

-  The Primary Combustion Air temperature is acquired during the Autozero phase. 

-  Insert the TcK connector of the gas sampling probe before starting the instrument and let it draw clean 

air, in order to get a correct temperature value. If the connector is not plugged in, the temperature will not 

be acquired. 

-  Storing of the Ambient Temperature: 

  Condensing Boilers /Furnaces:  

  Upon  Completion of the  Autozero in fresh  outdoor air, Press the KEEP button to  store  the  outdoor temperature 

being used for primary air 

  Atmospheric Boilers / Furnaces:   

  Upon  Completion  of  the  Autozero  in  fresh  outdoor  air,  move  the  E500  INDOORS  to  the  exact  location  of  the 

appliance  and  once  the  temperature  stabilizes  to  the  INDOOR  temperature,  Press  the  KEEP  button  to  store  the 

indoor temperature being used for primary air 

15/06/18  10:00 

PRINT 

 AutoZero in 

Fresh Air  
T: 76.8°F 

Keep 

Repeat 

         

15/06/18  10:00 

AUTOZERO 

 

 

60 

AutoZero in Fresh 

Air  

WARNING! 

-  Be  sure  all  connections  are  tight  to  assure  accurate 

sampling. 

-  It  is  important  that  water  trap  &  filter  be  installed 

vertically  to  prevent  damage  from  moisture  &  particles 

to sensors. 

-  When  conducting  measurements,  the  water  trap/filter 

assembly MUST be in a VERTICAL position. 

-  When  testing  is  completed,  always  drain  the  water  trap 

with any condensation (after EACH test!). 

To measure the draft proceed as follows: 

-  Connect  the  probe  pressure  input  hose 

to the instrument P+ input. 

   

-  Before  ZEROing  the  Pressure/Draft 

sensor,  please  be  sure  to  remove  the  gas  probe 

from the stack first  

- Upon completion  of the  Zeroing of the  Draft Sensor 

insert the probe in the stack to measure the Draft. 

It  is  compulsory  to  perform  the  autozero 

in  clean  air,  so  that  the  environment  CO 

measurement  results  correct.  It  is 

advisable  to  turn  on  the  instrument  and 

wait  for  the  autozero  completion  outside 

the  area  where  the  test  is  being 

performed. 

15/06/18  10:00 

DRAFT 

  Inlet           P+ 
  Draft       0.20 h 
► Zero sensor 

Save 

Print 

         

15/06/18  10:00 

AUTOZERO 

 

 

 

Zero sensor 

15/06/18  10:00 

DRAFT 

  Inlet           P+ 
  Draft       0.01 h 
► Zero sensor 

Save 

Print 

Save 

15/06/18  10:00 

CO AIR 

CO             412 p 
CO Max         413 p 

Save 

Print 

Save 

15/06/18  10:00 

PRINT 

 CO air 
►Copy number      1 
 Printer         IR 
 Mode          fast  
 QR Code        OFF 
 Pairing BT

 

 

Print 

         

15/06/18  10:00 

WARNING 

Printing 

in progress ... 

Print 

Print 

15/06/18  10:00 

DRAFT 

  Inlet           P+ 
  Draft       0.01 h 
► Zero sensor 

Save 

Print 

15/06/18  10:00 

CO AIR 

CO             412 p 
CO Max         413 p 

Save 

Print 

Print 

Reviews: