HUMIPORT 05 english
CAL2:
CAL2 (one-point calibration) is used to set an offset for sensor 2
(relative humidity). The offset is based on the lower adjustment point (11% RH).
The adjustment point must lie between 30% and 95% RH.
The offset value appears at the top of the display. The limit values are +/- 10% RH.
CAL2 can only be selected with the unit % RH.
The selection is made using
and
; and entered with .
To obtain the factory calibration, set the offset to 0.0.