User Manual EE360
Moisture in Oil Sensor | 29
6.5 Status Information
i
The status information shows all actual EE360 settings.
Fig. 27 Status information
6.6 Buzzer ON / OFF
Icon
Function
Buzzer ON
Buzzer OFF
Tab. 11
Explanation of the icons
6.7 Error Indication
When an error occurs, the error indication shows the error code.
7 Maintenance and Service
7.1 Self Diagnosis and Error Messages
7.1.1. Error Messages via Display
Error Description
Error Code (Display) Error Category Recommended Action
Voltage out short circuit - output 1 only*
1.1
1
Check the wiring of the outputs
Voltage out short circuit - output 2 only*
1.2
Voltage out short circuit - both outputs*
1.3
Current loop open - output 1 only
2.1
Check the wiring of the outputs
Current loop open - output 2 only
2.2
Current loop open - both outputs
2.3
RH sensor polluted
3.x
Clean the sensor
Hardware error
5.x
2
Return the faulty unit to E+E
for service
6.x
8.x
Temperature measurement failure
7.x
Humidity measurement failure
9.x
10.x
* not available with 0 - 1 V output
Tab. 12
Overview of error codes