Moisture in Oil Sensor | 26
6.1 Configuration Interface
Factory setup can be changed by using a USB configuration cable (USB-C to USB-A, accessory order code
HA010327) and the PCS10 Product Configuration Software. The EE360 is powered by the PC via the USB
interface, no additional power supply shall be applied.
The scaling of the analogue outputs, the settings of the alarm module, the digital settings can be changed and a T
and RH adjustment can be performed.
In addition, it is possible to enable or disable the NAMUR error indication (factory setting: disabled).
The PCS10 offers a convenient graphical user interface for the moisture in oil sensor. To use the software for
performing adjustments and changes in settings, please proceed as follows:
1.
Download the PCS10 Product Configuration Software from
2.
Connect the EE360 to the PC using the Modbus configuration adapter.
3.
Start the PCS10 software.
4.
Follow the instructions on the PCS10 opening page for scanning the ports and identifying the connected
device.
5.
Click on the desired setup or adjustment mode from the main PCS10 menu on the left and follow the online
instructions of the PCS10.
Blind cover
USB cable
Fig. 21 Micro USB serial interface
PCB
LED 1
LED 2
LED 4
LED 3
Current output
Error LED
error LED on = error category 1
error LED flashes = error category 2
Voltage output
power supply/
main board
Fig. 22 Status LEDs are located at the USB port
NOTICE
Ensure that the device is only powered by the USB interface during firmware update, otherwise the update
may fail.