BA_EE310_e // V1.5 // Technical data subject to change // 194474
Operation manual
EE310
Humidity/Temperature
Transmitter
Page 1: ...BA_EE310_e V1 5 Technical data subject to change 194474 Operation manual EE310 Humidity Temperature Transmitter...
Page 2: ...been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interferenc...
Page 3: ...2 3 4 Installation of the probe with ball valve set 13 4 Optional modules 15 4 1 Alarm module option AM2 15 4 2 Power supply module option AM3 17 4 3 Modbus RTU module option J3 17 4 4 Pluggable probe...
Page 4: ...hall be handled as such Do not touch the sensing elements inside the sensing head The replacement of the filter cap shall be done with utmost care not to touch the sensors Installation electrical conn...
Page 5: ...ultaneously and offers extensive error diagnostics The integrated data logging function saves all measured and calcu lated values to the internal memory The data can be displayed as graph directly on...
Page 6: ...l modules Fig 2 Modular enclosure 2 2 Models EE310 is available for wall or duct mount as well as with various remote probes The remote probes can be employed up to 180 C 356 F and 20 bar 300 psi T1 W...
Page 7: ...ameter 4 2 mm 0 2 not included in the scope of supply 54 2 13 104 4 09 71 2 80 5 1 0 20 166 6 54 180 7 09 153 6 02 54 2 13 14 0 55 166 6 54 5 1 0 20 7 0 28 Fig 4 Dimensions and mounting template of po...
Page 8: ...nds analogue range and scaling are freely selectable All settings can be performed via display and push buttons or using the EE PCS Product Configuration Software Error indication The analogue outputs...
Page 9: ...GND RS485 shielded RS485 A D n c Modbus RTU n c 2 1 4 3 5 RS485 B D GND RS485 shielded RS485 A D n c n c 2 1 4 3 5 RS485 B D GND RS485 shielded RS485 A D n c Modbus RTU n c Fig 7 Plug options for EE3...
Page 10: ...122 F with display The transmitter shall be mounted with the probe pointing downwards 166 6 54 54 2 13 Fig 8 Mounting of model EE310 T1 3 3 2 T2 Duct mounting EE310 T2 are duct mount transmitters For...
Page 11: ...sure ment to avoid T gradients along the probe In case of large T difference between the two sides of the separation wall it is highly recommended to insert the probe completely up to the cable outlet...
Page 12: ...0 T10 may be performed by qualified staff only Special attention shall be paid to the correct installation of the probe into the process In case of inappropriate installation there is the risk for the...
Page 13: ...wo metal sealing rings see Fig 13 Installation of the probe by utilizing the ball valve set shall be replaced every time prior to re installing the probe The temperature during probe installation may...
Page 14: ...the ball valve set Removing the probe Hold firmly the probe to avoid it being suddenly expulsed when releasing the lock not May sure nut to bend and damage the probe cable Loosen slowly the lock nut...
Page 15: ...ilable including hysteresis window and error indication NO NC C NC 6 NO 4 C 5 NC 3 NO 1 C 2 NO NC C Standard Inverted REL 1 REL 2 RELAY 2 RELAY 1 Fig 14 Alarm module option AM2 When error indication m...
Page 16: ...itching point Switching point Hysteresis Standard Inverted Normal state Switch window Measurand Switching point low Switching point Hysteresis Switching point high Switching point Hysteresis Standard...
Page 17: ...y by the E E Service National regulations for installation shall be observed 4 3 Modbus RTU module option J3 Instructions for Modbus Protocol Setup please see Application Note AN0103 www epluse com EE...
Page 18: ...t temperature C 32 bit float 31107 452 Dew point temperature F 32 bit float 31131 46A Frost point Dew point temperature C 32 bit float 31133 46C Frost point Dew point temperature F 32 bit float 31113...
Page 19: ...on Software or via display and push buttons Working pressure Probe replacement Parameters Humidity coefficient Resistance R0 Capacity C76 Capacity gain Capacity offset Temperature coefficient TK Resis...
Page 20: ...Remove the blind cover and plug in the USB cable PCB LED 1 LED 2 LED 4 LED 3 Current output Error LED error LED on error category 1 error LED flashes error category 2 Voltage output power supply main...
Page 21: ...min 15 min 30 min 1 h 5 h 12 h The data logger menu is also used to select the logged data to be shown as a graph as well as for scaling the graph Menu Sampling rate Data logging Axis 1 Measurand Axis...
Page 22: ...sition Fig 24 Data logging Each point in the graph represents a logged value The points are connected by a linear interpolation One point The data logging continues even when the data memory is full n...
Page 23: ...ds scaling error indication relay configuration mode set points state adjustment 1 and 2 point humidity temperature adjustment reset to factory adjustment calibration status configuration of Modbus da...
Page 24: ...ble with 0 1 V output Tab 2 Overview of error codes Error category 1 non critical error can be solved by the user The display blinks and the buzzer beeps every 10 seconds The red status LED flashes Er...
Page 25: ...ve humidity sensing element can be replaced by the user against an E E replacement sensor see chapter 8 Replacement parts Accessories Sensor replacement procedure Switch off supply voltage Remove the...
Page 26: ...er Depending on the level of accuracy required the humidity reference can be Humidity Calibrator e g Humor 20 please see www epluse com humor20 Handheld device e g Omniport30 please see www epluse com...
Page 27: ...iguration changes see user guide at www epluse com relabeling all versions Mating plug for integrated power supply AM3 Mating plug RKC 5 7 AM3 E4 E7 E12 Mating plug RSC 5 7 2 pcs for option E12 E5 E7...
Page 28: ...0 60 Hz with option AM3 3 Current consumption 2x voltage output for 24V DC AC typ 40 mA 2x current output typ 80 mA Pressure range for pressure tight probe 0 01 20 bar 0 15 300 psi Probe material stai...
Page 29: ...1 Measurand Axis 2 Measurand Scale maximum value value Scale minimum OFF 1 s 10 s 1 min 5 min 15 min 30 min 1 h 5 h 12 h Delete saved values OFF Temperature C Temperature F Temperature K Rel humidity...
Page 30: ...t K Frost point K Volume concentr ppm Mode Current Voltage ON Output number Output 1 Output 2 OFF Measurand min max value OK value Cancel Analog range Scaling Error indication OFF Temperature C Temper...
Page 31: ...ing system Working pressure Probe replacement Parameters Humidity coefficient Resistance R0 Capacity C76 Capacity gain Capacity offset Temperature coefficient TK Resistance Offset Code Daylight saving...
Page 32: ...steresis Standard Inverted Normal state Error indication Standard Inverted Normal state OFF Temperature C Temperature F Temperature K Rel humidity Vap p pressure mbar Vap p pressure psi Dew point C De...
Page 33: ...33 Operating instructions for EE310 Humidity Temperature Transmitter 11 Notes...
Page 34: ...34 Operating instructions for EE310 Humidity Temperature Transmitter...
Page 35: ...35 Operating instructions for EE310 Humidity Temperature Transmitter...
Page 36: ...cn BEIJING Tel 86 10 84992361 info epluse cn SHANGHAI Tel 86 21 61176129 info epluse cn E E GERMANY www epluse de Tel 49 6172 13881 0 info epluse de E E FRANCE www epluse fr Tel 33 4 7472 35 82 info...