E+E Elektronik EE074 User Manual Download Page 6

6

User Manual EE074 Temperature Probe with Modbus RTU

4.2  Duct Mount

The probe can be installed also into a duct using the plastics flange 

HA401101 (not included in the scope of supply, see data sheet 

“Accessories”).

4.3  Immersion Well Mount

The probe can be installed into a pressured liquid system using an  

ISO or NPT ½“ immersion well HA400101-04 or HA4000211-14  

(not included in the scope of supply, see data sheet “Accessories”).

Please note:

 For a good thermal connection between the probe and the immersion well, the use of a 

thermal paste is recommended.

The EE074 can be used up to 15 bar (218 psi) with an E+E brass immersion well and with an E+E 

stainless steel immersion well up to 25 bar (363 psi).

Please observe following maximum flow speed limitations for the E+E immersion wells:

Length (L)

50 mm 

(1.97“)

100 mm

 (3.94“)

135 mm

 (5.31“)

285 mm

 (11.22“)

Brass

26 m/s 

(5118 ft/min)

12 m/s 

(2362 ft/min)

6 m/s 

(1181 ft/min)

1 m/s 

(197 ft/min)

Stainless steel

29 m/s 

(5118 ft/min)

15 m/s 

(2953 ft/min)

9 m/s 

(1771 ft/min)

2 m/s 

(394 ft/min)

4.4  Recommendations for Accurate Temperature Measurement

Air Temperature Measurement

In air, best accuracy is achieved with an air flow of min. 0.2 m/s surrounding the sample.
In case the EE074 probe is not entirely located in the environment to be monitored, large temperature 

differences along the probe will lead to temperature gradients. These will have an influence on the 

accuracy. Therefore, it is of paramount importance to minimize the temperature gradients. The biggest 

part of the probe shall be located in the target environment and the rest shall be thermally well insula-

ted.

Air and Liquids Temperature Measurement down to -70 °C (-94 °F)

The front part of the 305 mm (12.07″) probe may be immersed up to 100 mm in a medium with a 

temperature of down to -70 °C (-94 °F). The temperature of the other parts, especially those containing 

the electronics, shall not fall below -40 °C (-40 °F). The insulation design depends on the measurement 

task. In good thermal conductors like in liquids or dry block calibrators, the insulation can be smaller in 

order to achieve a good measurement accuracy. In contrast, for media with low thermal conductivity like 

air, the insulation must cover at least 200 mm (7.87″) of the remaining probe.

Summary of Contents for EE074

Page 1: ...BA_EE074_e V1 0 technical data are subject to change User Manual EE074 Temperature Probe with Modbus RTU...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...5 4 2 Duct Mount 6 4 3 Immersion Well Mount 6 4 4 Recommendations for Accurate Temperature Measurement 6 5 Modbus Map 7 5 1 Modbus Setup 7 5 2 Modbus Register Map 7 5 3 Modbus RTU Example 8 6 Maintena...

Page 4: ...facturing conditions has been thoroughly tested and has left the factory fulfilling all safety criteria The manufac turer has taken all precautions to ensure safe operation of the device The user must...

Page 5: ...in mm inch 82 3 23 L 12 0 47 6 0 24 L in mm L in Inch 70 2 81 155 6 16 305 12 07 3 2 Connectivity 4 Installation 4 1 Wall and Ceiling Mount Best measurement performance is achieved when the entire pr...

Page 6: ...4 Recommendations for Accurate Temperature Measurement Air Temperature Measurement In air best accuracy is achieved with an air flow of min 0 2 m s surrounding the sample In case the EE074 probe is no...

Page 7: ...dbus protocol settings see Application Note Modbus AN0103 available on www epluse com EE074 INFO read register Parameter Register number1 DEC Protocol address2 HEX Read only register function code 0x0...

Page 8: ...any special maintenance nevertheless for high accurate measurements especially over wide T ranges it is recommended to calibrate the probe every 12 months If needed the enclosure may be cleaned and t...

Page 9: ...ed setup or adjustment mode from the main EE PCS menu on the left and follow the online instructions of the EE PCS PC HA011018 6 4 T Adjustment and Calibration 6 4 1 Temperature Adjustment and Calibra...

Page 10: ...ectronics 40 80 C 40 176 F 0 100 RH probe 70 155 mm 40 80 C 40 176 F 0 100 RH probe 305 mm 70 105 C 94 221 F 0 100 RH Storage conditions 40 80 C 40 176 F 0 90 RH Immersion well Material brass nickel p...

Page 11: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Reviews: