Dytran 3224C Operating Manual Download Page 3

REV C ECN 10389 09/24/13 

 

3

OPERATION 

 
 

To operate Model 3224C, it is necessary to 

use an electrostatic charge amplifier to convert the 
charge mode output signal from the accel. to a low 
impedance voltage mode signal which may then be 
fed directly to the readout instrument(s). 
 
 

A recommended charge amplifier for field use 

is the Dytran Series 4751B. These instruments convert 
the 3224C to Low Impedance Voltage Mode (LIVM) 
operation. They are mounted in-line between the 
3224C and the LIVM power unit and convert the pC/G 
signal to mV/G. These miniature charge amplifiers are 
available in four models with fixed sensitivities of 50, 
20, 10 and 1 mV/pC. This yields nominal system 
sensitivities of 30, 30, 12 and 0.6 mV/G, respectively. 
 

 
For laboratory use, 3224C may be used with 

any vibration type laboratory charge amplifiers and 
even with most electrostatic types. Consult the factory 
if there is a question about compatibility with available 
charge amplifiers. 

 

 

The cable supplied with Model 3224C is 

terminated in a jack type (male thread) 10-32 coaxial 
connector. The recommended extension cables are 
the Dytran low noise coaxial cables, model 6013AXX 
(10-32 plug to 10-32 plug) or the model 6053 (10-32 
plug to BNC plug). Both cables mate directly with the 
cable attached the 3224C. Other cables are available 
if either of these does not meet your requirements. 
Contact the Dytran sales department for help in 
selecting the best cable for your requirement. 
 
 

The polarity convention of Model 3224C is 

negative charge output for acceleration toward the top 
of the unit. This is so because most charge amplifiers 
are inverting amplifiers, hence the resultant system 
signal will be positive with the above input convention. 
 

UNMOUNTING THE ACCELEROMETER 
 
 

In order to "unmount" the Model 3224C, use 

the Model 6725 tool. The larger slotted end of the tool 
is used for this purpose. Slip the tool over the 
accelerometer body from the rounded end and gently 
rotate the tool in either direction until the adhesive 
shears and the instrument is released. 
 
 

Do not use 

pliers, wrenches and other tools to 

remove the instrument, as these are certain to mar or 
otherwise damage the unit. 

 
 
 

MAINTENANCE AND REPAIR 

 
 

The only maintenance necessary is to keep 

cables and connections clean and free from moisture. 
All charge mode systems are considered high 
impedance systems and as such are susceptible to 
moisture which degrades insulation resistance. 
 
 

Should a problem arise with the accelerometer 

or cable, contact the factory for assistance in trouble 
shooting or returning the instrument for evaluation 
and/or repair. Do not send the instrument back without 
first calling the factory to obtain a Returned Material 
Authorization (RMA) number. 

Reviews: