Installation, Operation and Service Manual
LKAT Plus EMC
© 2018 DynAmp, LLC
Page 3
047331 A
2. PRODUCT DESCRIPTION
2.1
SYSTEM OVERVIEW
The DynAmp LKAT Plus system is designed to accurately measure AC or DC current flowing
in a bus bar. Each system consists of a Measuring Head, a Metering Unit, and
multi-conductor cables.
The LKAT Plus Metering Unit consists of a power supply pc board and Main pc board
mounted together on a steel U-chassis. Input, output, and power connections are made from
outside the enclosure using panel-mount IP65 rated military-style circular connectors. Wiring
harnesses within the Metering Unit are terminated using 3.5mm and 5.08mm “Eurostyle”
connectors.
AVAILABLE ACCESSORIES
•
Summing / Totalizing multiple LKAT systems
A dedicated external module is available for totalizing/ summing up to 10 LKAT measurement
signals. This is typically used to provide a true ‘total’ current signal when multiple rectifiers
( with individual LKAT systems ) are connected in parallel to provide a higher total current.
•
LKAT-MUT Functional Test
This hand-held test set allows users to verify system operation and scaling as well as test
relay trip points and output signals in the field.
•
LKAT Metering Unit Tester ( LKAT MUT )
The LKAT MUT is a portable tester used to verify the functionality of Metering Units for all
versions of the LKAT family of products.
Adapters included with the LKAT MUT allow the tester to be connected directly and check
any LKAT Metering Unit electronics.
The functions and parameters checked by the LKAT MUT include:
•
Input / output sensitivity
•
Trip point scaling
•
Trip relay contacts
•
Integrity of input / output harnesses and interconnections on Metering Unit
The tools / documentation required to use the LKAT Metering Unit Tester include :
•
4-1/2 digit DMM with 0.05% basic DC accuracy ( user provided )
•
DMM Test probes ( user provided )
•
Banana to banana cables ( 1black / 1 red )
•
LKAT MUT Product Manual
More information on the LKAT MUT can be found at
DynAmp.