All rights reserved. Reproduction or issue to third parties in any form whatsoever is not permitted without written authority from the proprietors.
5
2.3) Specifications
Table 1: Specifications datasheet
Full scale
acceceleration
(g)
3002SE
±2
3004SE
±4
3008SE
±8
3010SE
±10
3020SE
±20
3040DE
±40
3050SE
±50
3100SE
±100
3200SE
±200
3500SE
±500
Frequency
range (±3dB)
(Hz)
1,500
1,500
1,500
1,500
1,500
1,500
3,000
3,000
3,000
3,000
Non-linearity
(full scale)
(%)
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
Noise (in band)
(μg/
√Hz)
25
25
25
80
80
110
130
220
550
1,200
Scale factor
(nominal)
(mV/g)
400
200
100
80
40
20
40
20
10
4
Shock
survivability
(g)
5,000
5,000
5,000
5,000
5,000
5,000
5,000
6,000
6,000
6,000
Environmental
Table 2 Environmental specifications datasheet
Protection Level
IP 68
Operating Voltage
6 V
–
20 V
Operating Temperature
-40 °C to +100 °C
Operating Current Consumption mA
21 mA
Isolation
Case isolated