MK5 Eco Panel overview 

Introduction  

Welcome to the Dynaflow ® MK5 Eco Control Panel. This panel has been designed by Dynaflow 
specifically for the Dynaflow Fume Cupboard. The MK5 Panel has been “Product Certified” ISO 
Product Certification to compliance to AS/NZS 2243.8 2014. The MK% has been designed to provide 
simple “Function” changes and ease of “Plug and Play” accessories to allow for quick upgrade and 
replacement options. 
Please find and overview of the MK5 Eco Control Panel below. 

Panel Layout 

The Dynasafe® Control System MK5 Fume cupboard controller has 4 key aspects 
represented on the front panel. 
 

 

A Graphical LCD display which provide the bulk of the user feedback information. 

 

5x press buttons which allow user interaction. 

 

 USB port. 

 

 Emergency Stop button.

 

Panel functions 

1.

 

Air flow indicator -

 indicates when air flow within the cupboard and duct work 

meets acceptable requirements 

2.

 

Display -

 indicates cupboard status to the operator. Shows time, current mode, 

status information and key functions. Refer display breakdown for more information. 

3.

 

Emergency Isolator

 - used to isolate the cupboard from electrical and gas services in 

the event of an emergency. 

4.

 

Supply Fault -

 indicates Mains failure, will flash red should Mains supply fail. The 

controller is then operating on batteries. 

5.

 

Function switch

 - used by the operator to enter and exit setup and configuration 

options. 

6.

 

USB Port -

 This port provides USB connection to external computers. It is primarily 

used for test and commissioning of the cupboard. It provides no user data. 

7.

 

Light switch -

 controls the cupboard lighting. 

8.

 

Fan switch -

 used to start and stop the cupboard. 

9.

 

Boost switch -

 if the cupboard is fitted with Syncroflow® then can be used to boost 

airflow. 

10.

 

Mute Switch

 - used to mute buzzer during alarm states. Please note that should a 

new alarm occur then the mute will be removed from the alarm. Once the alarm has 
been corrected then the buzzer will turn off automatically. 

Summary of Contents for MK5 Eco

Page 1: ...able requirements 2 Display indicates cupboard status to the operator Shows time current mode status information and key functions Refer display breakdown for more information 3 Emergency Isolator used to isolate the cupboard from electrical and gas services in the event of an emergency 4 Supply Fault indicates Mains failure will flash red should Mains supply fail The controller is then operating ...

Page 2: ...ncroflow and buzzer mute Together with the function button they have alternate functions that are used to setup and configure the fume cupboard Light Control s the cupboard lighting Toggle function press to turn on and press to turn off Will operates in any of the cupboard s operational modes Fan Control s the cupboard operational modes Toggle function press to start cupboard Starts 60 Second PRE ...

Page 3: ...by displaying RUNNING on the controller display Should an alarm situation arise during RUNNING mode then the fume cupboard will turn off services and sound the alarm and provide indication as to cause of the alarm Once the problem has been resolved then the fume cupboard will restart in PRE PURGE mode Post purge When the handling process is complete the fume cupboard needs to be shutdown Press FAN...

Page 4: ... specifications Emergency Isolator Located on the front panel this switch will cause the fume cupboard to shut down and isolate all electrical and gas services To restore operation twist and release the emergency isolator switch Mains Failure Indicates electrical power failure Batteries will allow operation of the controller for minimum 20 minutes from power failure Low Battery This will occur if ...

Page 5: ...r the keep the sash in a safe position but reduces the amount of exhaust air being extracted from the Fume Cupboard thus reducing the huge energy consumption required in the laboratory make up air system Scrubber PH Monitoring When enabled the MK5 controller monitors external PH input via 4 20ma current loop Scrubber System OPTION The MK5 Eco Panel has available 40 programmable functions available...

Page 6: ...h Warning Dual Sash Green Sash BMS Interface Greenlight Service Meter Pump 1 Pump 2 Valve 1 Valve 2 VSD Direct Fan Disable AUTO Sash Services Process Timers Prepurge and Postpurge Timers Green Sash Commissioning Service Call System Mode BMS Interface Waste Water Dump Sash Close Time Config Sash PH Air VSD Settings Sash Sensor Air Sensor PH Sensor Serial No Change Password BLight Test Relay Outputs...

Page 7: ...relays 18 Alarm Relay General alarm relays 19 Sash Input Require to calibrate sash s 20 Air Sensor Variable Hot wire airflow sensor N A 21 Commissioning Activate commissioning setup 22 Auto Start Auto start after power failure Allows Services 23 RS485 BMS interface High Level N A 24 Remote Panel Additional control panel D S Fume Cupboards 25 PH Meter Monitor PH Control alarms only 26 Air Con Start...