ymax ZIP™ Shutter User Guide
Electrical Interconnect
Figure 12. ZIP Shutter Electrical Interconnect Diagram
1.
Connect Shielded Cable (4) from J2 on rear of Power Supply (2) to J3 on rear of Reflector Housing (3).
2.
Connect Cable (5) from J4 on rear of Reflector (3) to 8-Pin Connector on rear of Shutter (1).
3.
Connect Plug (6) into J5 on rear of Reflector (3).
NOTE:
Shutter will not operate unless Plug is installed.
4.
Install Foot Switch to 4-Pin Connector on rear of Shutter (7).
5.
Connect Power Cord to J1 on rear of Power Supply (2).
Dymax
Flood Lamp Electrical Interconnect With Retrofit
To retrofit Shutters for use with older Dymax "Blue Series" Flood Lamps, please contact the Dymax
Application Engineering or Equipment Group.
Operation
1.
Have the ZIP Shutter in the closed position.
2.
Open the Access Door to the Workstation or Dymax
Light Shield and place the object to be cured on
the work surface (preferably in line with the center of the Shutter).
NOTE:
The ZIP Shutter will close automatically if the Light Shield has an Interlock Kit (PN 41166)
installed and the Access Door is opened.
3.
Adjust the distance from the bottom plane of the
ZIP Shutter
to the surface to be cured.
4.
Close the Access Door.
5.
Set the
exposure (ZIP Shutter “Open Time") by adj/
- the Timer located next to the Actuate
Button on the Front Panel.
6.
Open the ZIP Shutter using the most convenient method for you by using Push Actuator Button or
Foot Switch.