DISPLAY MENU: diS – MENU

With  the  gage  reading  a  numerical  value  (home  position),  press  and  hold  the

MENU

key until the Security MENU is displayed. Press the 

q

key until the LCD

shows 

diS – MENU

. Press the 

E

key to show the Display SUB MENUS.

FILTER TYPE SETTING

With the LCD reading 

FILt – OFF

, press the 

E

key; the display will blink. Press the

q

key to change the filter type then press the 

E

key to save the desired filter type.

FILt – OFF:

Filter type = OFF. The filter function is disabled.

FILt – HIGH:

Filter type = HIGH. The display blinks when pressure is greater than

the filter point.

FILt – LOW:

Filter type - LOW. The display blinks when pressure is less than the

filter point.

FILTER POINT SETTING

With the LCD reading 

0.00 – SPPT

, press the 

E

key; the display will blink. Press

the 

q

key to change the filter point then press the 

E

key to save the desired filter

point.

0.00 – SPPT:

Filter point = 0.00. The filter point may be set to anywhere within the

range of the instrument.

DAMPING SETTING

With the LCD reading 

1 – DAMP

, press the 

E

key; the display will blink. Press the

q

key  to  change  the  damping  level  then  press  the 

E

key  to  save  the  desired

damping level.

1 – DAMP:

Damping level = 1. The damping level can be adjusted from 1 to 15.

Damping stabilizes the display from instabilities due to things such as vibration and
excessive  pressure  fluctuations.  The  damping  function  adjusts  the  amount  of
readings that are averaged for each display update.

DISPLAY UPDATE SETTING

With the LCD reading 

rAtE – NORM

, press the 

E

key; the display will blink. Press

the 

q

key  to  change  the  display  update  rate  then  press  the 

E

key  to  save  the

desired display update rate.

rAtE – NORM:

Display update = Normal (1 second).

The gage reads the process pressure and updates the LCD every second.

rAtE – 10:

Display update = 10 minutes.

The gage reads the process pressure and updates the LCD every 10 minutes.

rAtE – KEY (On-Touch mode):

Display update is disabled.

The gage reads the process pressure and updates the LCD whenever the 

key

is pressed. If the 

p

key is released, the LCD will hold and display the last pressure

reading.

NOTES:

1. Depending on the Display Update setting, the battery life is shown below:

-150 hours (typical) if Display Update is set for “Normal”  1 second update. rAtE-   

NORM
-9 months (typical) if Display Update is set for 10 minutes. rAtE-10
-1.5 years (typical) if Display Update is disabled. rAtE-KEY

2. If the Display Update is set for 10 minutes or disabled (On-Touch mode), the
process pressure value can be read instantly by pressing and holding the 

p

key

on the front panel. Also the LCD will automatically show “ALAr” if the filter point has
been exceeded.

PEAK AND VALLEY SETTING
100.0 – PEAK: 

Peak value = 100.0

The peak feature stores the highest pressure reading the instrument has measured
since the last reset or power up. At power up 

PEAK

is reset to the present pressure

reading. To  manually  reset  the 

PEAK

value,  press  the 

E

key  while  in  the 

PEAK

SUB MENU.

0.0 – VALY: 

Valley value = 0.0

The valley feature stores the lowest pressure reading the instrument has measured
since the last reset or power up. At power up 

VALY

is reset to the present pressure

reading. To manually reset the 

VALY

value, press the 

E

key while in the 

VALY 

SUB

MENU.

ADVANCED MENU: AdU – MENU

With  the  gage  reading  pressure  (home  position),  press  and  hold  the 

MENU

key

until the Security MENU is displayed. Press the 

q

key until the display shows the

AdU – MENU

. Press the 

E

key to show the advanced function SUB MENUS.

Auto ZERO

: auto-zero

NOTE: For  accurate  calibration,  DO NOT  apply  any  pressure  when
performing this function.

With the display reading 

Auto ZERO

, release pressure to Zero then press the 

E

key; the display will blink. Press the 

E

key again to complete the auto-zero.

CAL – SPAN:

full-scale calibration

With the display reading 

CAL – FS

, apply full-scale pressure then press the 

E

key;

the  display  will  blink.  Press  the 

E

key  again  to  save  the  full-scale  calibration  or

press the 

MENU

key to cancel the calibration.

MAINTENANCE

This model is not field-serviceable and should be returned if repair is needed (field
repair should not be attempted and may void warranty). Be sure to include a brief
description of the problem plus any relevant application notes. Contact customer
service to receive a return goods authorization number before shipping.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Series DM-1000

Page 1: ...ECIFICATIONS Service Air and non combustible compatible gases Wetted Materials Consult factory Housing Materials Glass filled plastic Accuracy 1 F S including linearity hysteresis and repeatability 2 F S for ranges 1 w c and below Temperature Limits 0 to 140 F 18 to 60 C Compensated Temperature Limits 32 to 122 F 0 to 50 C Long Term Stability 1 F S per year Thermal Effect 0 05 F S F typ 0 10 F S F...

Page 2: ...tallation The unit is shipped with a separate 9V alkaline battery Remove the unit s top back cover do not remove bottom cover Connect the 9V battery to the battery holder as shown in Figure 1 Replace top back cover and tighten screws to 2 25 in lbs External Power Supply Installation Remove the unit s back cover then connect the external power supply to the terminal block as shown in Figure 1 OPERA...

Page 3: ...Menu Map www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 4: ...ation SUB MENUS for Series DM 1100 is PrES INWC Pressure Unit PRESSURE SUB MENU PRES INWC PrES INWC Pressure inches of water column With the LCD reading PrES INWC press the E key the display will blink Press the q key to change the pressure unit then press the E key to save the desired unit For pressure measurement the following units are available INWC Inches of water column MBAR Millibar PSI Pou...

Page 5: ...1 Depending on the Display Update setting the battery life is shown below 150 hours typical if Display Update is set for Normal 1 second update rAtE NORM 9 months typical if Display Update is set for 10 minutes rAtE 10 1 5 years typical if Display Update is disabled rAtE KEY 2 If the Display Update is set for 10 minutes or disabled On Touch mode the process pressure value can be read instantly by ...

Page 6: ...LE DIMENSION FOR DIGIMAG GAUGE Ø4 9 16 115 89 1 25 44 2X30 2xØ13 32 10 32 THRU HOLES FOR PRESSURE PORTS 1 51 64 45 72 3 15 32 88 12 3 SURFACE MOUNTING HOLE LOCATIONS 13 32 10 16 Ø5 8 15 88 THRU HOLE FOR CONDUIT CONNECTION OUTSIDE DIAMETER OF A 286 FLANGE 1 1 4 31 75 3 FLANGE MOUNTING HOLE LOCATIONS Ø5 17 32 140 49 BOLT CIRCLE MOUNTING HOLE DIMENSION FOR A 286 FLANGE Ø5 1 8 130 18 www GlobalTestSup...

Reviews: