DISPLAY MENU: diS – MENU
With the gage reading a numerical value (home position), press and hold the
MENU
key until the Security MENU is displayed. Press the
q
key until the LCD
shows
diS – MENU
. Press the
E
key to show the Display SUB MENUS.
FILTER TYPE SETTING
With the LCD reading
FILt – OFF
, press the
E
key; the display will blink. Press the
q
key to change the filter type then press the
E
key to save the desired filter type.
FILt – OFF:
Filter type = OFF. The filter function is disabled.
FILt – HIGH:
Filter type = HIGH. The display blinks when pressure is greater than
the filter point.
FILt – LOW:
Filter type - LOW. The display blinks when pressure is less than the
filter point.
FILTER POINT SETTING
With the LCD reading
0.00 – SPPT
, press the
E
key; the display will blink. Press
the
q
key to change the filter point then press the
E
key to save the desired filter
point.
0.00 – SPPT:
Filter point = 0.00. The filter point may be set to anywhere within the
range of the instrument.
DAMPING SETTING
With the LCD reading
1 – DAMP
, press the
E
key; the display will blink. Press the
q
key to change the damping level then press the
E
key to save the desired
damping level.
1 – DAMP:
Damping level = 1. The damping level can be adjusted from 1 to 15.
Damping stabilizes the display from instabilities due to things such as vibration and
excessive pressure fluctuations. The damping function adjusts the amount of
readings that are averaged for each display update.
DISPLAY UPDATE SETTING
With the LCD reading
rAtE – NORM
, press the
E
key; the display will blink. Press
the
q
key to change the display update rate then press the
E
key to save the
desired display update rate.
rAtE – NORM:
Display update = Normal (1 second).
The gage reads the process pressure and updates the LCD every second.
rAtE – 10:
Display update = 10 minutes.
The gage reads the process pressure and updates the LCD every 10 minutes.
rAtE – KEY (On-Touch mode):
Display update is disabled.
The gage reads the process pressure and updates the LCD whenever the
p
key
is pressed. If the
p
key is released, the LCD will hold and display the last pressure
reading.
NOTES:
1. Depending on the Display Update setting, the battery life is shown below:
-150 hours (typical) if Display Update is set for “Normal” 1 second update. rAtE-
NORM
-9 months (typical) if Display Update is set for 10 minutes. rAtE-10
-1.5 years (typical) if Display Update is disabled. rAtE-KEY
2. If the Display Update is set for 10 minutes or disabled (On-Touch mode), the
process pressure value can be read instantly by pressing and holding the
p
key
on the front panel. Also the LCD will automatically show “ALAr” if the filter point has
been exceeded.
PEAK AND VALLEY SETTING
100.0 – PEAK:
Peak value = 100.0
The peak feature stores the highest pressure reading the instrument has measured
since the last reset or power up. At power up
PEAK
is reset to the present pressure
reading. To manually reset the
PEAK
value, press the
E
key while in the
PEAK
SUB MENU.
0.0 – VALY:
Valley value = 0.0
The valley feature stores the lowest pressure reading the instrument has measured
since the last reset or power up. At power up
VALY
is reset to the present pressure
reading. To manually reset the
VALY
value, press the
E
key while in the
VALY
SUB
MENU.
ADVANCED MENU: AdU – MENU
With the gage reading pressure (home position), press and hold the
MENU
key
until the Security MENU is displayed. Press the
q
key until the display shows the
AdU – MENU
. Press the
E
key to show the advanced function SUB MENUS.
Auto ZERO
: auto-zero
NOTE: For accurate calibration, DO NOT apply any pressure when
performing this function.
With the display reading
Auto ZERO
, release pressure to Zero then press the
E
key; the display will blink. Press the
E
key again to complete the auto-zero.
CAL – SPAN:
full-scale calibration
With the display reading
CAL – FS
, apply full-scale pressure then press the
E
key;
the display will blink. Press the
E
key again to save the full-scale calibration or
press the
MENU
key to cancel the calibration.
MAINTENANCE
This model is not field-serviceable and should be returned if repair is needed (field
repair should not be attempted and may void warranty). Be sure to include a brief
description of the problem plus any relevant application notes. Contact customer
service to receive a return goods authorization number before shipping.