Dwyer Instruments LOVE Iso Verter II SC478 Installation Instructions Manual Download Page 7

February, 2000

Page 7

949-0533

Specifications

Power Supply: SC478:  85 to 265 Vac/ Vac 50 to 400 Hz.

SCL478:  12 to 24 Vdc/Vac 50 to 400 Hz. ±20%

Isolation:

1500 VAC

Ambient Temperature  Range:

Operating:

0° to 55° C (32° to 131° F)

Storage:

-40° to +80°C (-40° to +176°F)

Humidity Conditions: 0 to 90% up to 40 °C non-condensing, 10 to 50%
at 55 °C non-condensing.
Linearity:

0.1%

Drift:

±0.02% per °C typical, ±0.05% maximum.

Maximum current output load:

600 ohms.

Maximum voltage output load:

20mA (500 ohms).

Input Type: Frequency.
Full scale range: 1 to 10,000 Hz.
Input impedence:

>50 Kohms.

Input Amplitude: 50mV to 150V RMS, overload protected to 180V RMS
Zero adjustment: 0-9998 Hz.
Span adjustment: 2-10,000 Hz.
Minimum span: Input<1,000 Hz = 0.2 Hz.  Input >1,000 kHz = 2 Hz.
Input Resolution:

Maximum calibrated input frequency <1,000 Hz = 0.01 Hz
Maximum calibrated input frequency >1,000 Hz = 0.1 Hz

Output Resolution: >7000 steps (approximately 13 bits).

Summary of Contents for LOVE Iso Verter II SC478

Page 1: ...Dwyer Instruments Incorporated PO Box 338 Michigan City IN 46361 0338 800 828 4588 219 879 8000 FAX 219 872 9057 www love controls com LOVE CONTROLS LOVE 1 Delta Park Blvd 12 Brampton ON L6T 5G1 Tel...

Page 2: ...INSTRUMENTS INC ALL RIGHTS RESERVED PRINTED IN U S A Contents DESCRIPTION 3 OUTPUT PROGRAMMING HARDWARE 3 MODE SELECTION SWITCH BANK SW 3 3 RANGE SELECTION SWITCH BANK SW 4 4 MOUNTING 4 WIRING 4 CALIB...

Page 3: ...uencies between 1000 Hz to 10 000 Hz the minimum span is 2 Hz This allows the input to be pro grammed for almost any desired span i e 10 to 15 Hz 3456 to 5678 Hz or 9998 to 10 000 Hz Output Selection...

Page 4: ...tional surface mounting kit is available from the factory P N 35DINADPTR To install hold the SC478 so that the front is higher than the rear Place the upper slot on the rear of the SC478 on the top ed...

Page 5: ...LED will light Note if the LEDs were flashing due to a previous error condition they will stop and allow calibration to continue 3 Input the MINIMUM desired frequency 4 Press the CAL switch the yellow...

Page 6: ...desired input value and then adjust the input frequency for the desired MINIMUM output value 4 Press the CAL switch The yellow OUT LED will turn off and the cali bration values will be stored in nonv...

Page 7: ...1 Drift 0 02 per C typical 0 05 maximum Maximum current output load 600 ohms Maximum voltage output load 20mA 500 ohms Input Type Frequency Full scale range 1 to 10 000 Hz Input impedence 50 Kohms Inp...

Page 8: ...rated PO Box 338 Michigan City IN 46361 0338 800 828 4588 219 879 8000 FAX 219 872 9057 www love controls com LOVE CONTROLS LOVE 1 Delta Park Blvd 12 Brampton ON L6T 5G1 Tel 905 457 6322 or 1 800 794...

Reviews: