Dwyer Instruments BCHP Series Installation And Operating Instructions Download Page 2

PRESSURE OPERATION
Step 1:

Check whether the change-over switch (4) has pressure selected.

(See label on neck of the pump) If pressure is not selected, use a small tool
to push the switch to the correct position. 

Step 2:

Make sure the pressure relief knob (3) is not completely closed.

Step 3: 

Turn the fine adjustment knob (2) counter-clockwise until it stops.

Then,  turn  the  knob  three  turns  clockwise  to  ensure  the  adjustment  is
adequate to increase or decrease the pressure.

Step 4:

Turn the pressure relief valve (3) until it closes.

Step 5:

Press down on the pump handle (6) until the approximate pressure

level has been reached. 

Step  6: 

Turn  the  fine  adjustment  valve  (2)  clockwise  to  increase  the

pressure or counter-clockwise to decrease the pressure. Do this until the
required pressure level has been reached on the reference instrument. 

Step 7: 

To reduce the pressure in the pump, turn the fine adjustment valve

(2) counter-clockwise.  If  more  pressure  still  needs  to  be  decreased,
carefully turn the pressure relief knob (3).

VACUUM OPERATION
Step 1: 

Check whether the change-over switch (4) has vacuum selected.

(See label on the neck of the pump) If vacuum is not selected, use a small
tool to push it to the correct position. 

Step 2: 

Make sure the pressure relief knob (3) is not completely closed.

Step 3: 

Turn the fine adjustment knob (2) clockwise until it stops.  Rotate

the knob counter clockwise three turns so that the vacuum pressure can be
increased or decreased.

Step 4:

Turn the pressure relief knob (3) until it closes.

Step 5: 

Press down on the pump handle (6) until the approximate vacuum

level has been reached. 

Step  6: 

Turn  the  fine  adjustment  valve  (2)  clockwise  to  decrease  the

vacuum  or  counter-clockwise  to  increase  the  vacuum.  Do  this  until  the
required pressure has been reached on the reference instrument. 

Step 7:

To reduce the vacuum in the pump, turn the fine adjustment valve

(2) counter-clockwise. If additional vacuum pressure needs to be relieved,
carefully open the pressure relief valve (3).

FAULT INVESTIGATION

1. If  the  pressure  or  vacuum  cannot  be  generated  correctly  or  if  the  set
pressure and vacuum does not stay stable, it may be caused by incorrect
positioning  or  sealing  of  the  gaskets.  Be  sure  any  adapters  on  the  test
instrument have been sufficiently tightened.

2. Before assuming there is a leak in the pump, check if the pressure relief
valve (3) is closed and if the change-over switch (4) is correctly positioned.

3. Do not apply any force to the operating elements of the calibration test
pump and do not connect an external pressure supply to the pump.

MAINTENANCE/REPAIR

Upon  final  installation  of  the  Series  BCHP,  no  routine  maintenance  is
required. The Series BCHP is not field serviceable and should be returned
if  repair  is  needed.  Field  repair  should  not  be  attempted  and  may  void
warranty.

WARRANTY/RETURN

Refer to “Terms and Conditions of Sales” in our catalog and on our website.
Contact customer service to receive a Return Goods Authorization number
before  shipping  the  product  back  for  repair.  Be  sure  to  include  a  brief
description of the problem plus any additional application notes.

Do not exceed a pressure of 650 psi (45 bar).

NOTICE

After increasing the pressure, the reading may slightly drop for
about  30  seconds,  due  to  thermodynamic  effects,  the  tube

connection, and the sealing gaskets. (If the pressure does not come to a standstill
within that time, check the ports to be sure they are sealed tightly.)

NOTICE

Do not exceed a vacuum of -26.5 in Hg (-0.9 bar.)

NOTICE

After increasing the vacuum, the reading may slightly increase
for  about  30  seconds,  caused  by  thermodynamic  effects,  the

tube  connection,  and  the  sealing  gaskets.  (If  the  vacuum  does  not  come  to  a
standstill within that time, check the ports to be sure they are sealed tightly. )

NOTICE

©Copyright 2014 Dwyer Instruments, Inc.

Printed in U.S.A. 7/14

FR# R6-444183-00

Do not move the change-over switch (4) when the test pump is
under operation.

NOTICE

Only  remove  the  reference  or  test  instrument  when  the  relief
valve (3) is open and no pressure is in the test pump.

NOTICE

Do not move the change-over switch (4) when the test pump is
under operation.

NOTICE

Only  remove  the  reference  or  test  instrument  when  the  relief
valve (3) is open and no vacuum is in the test pump.

NOTICE