background image

Bulletin TE-BCHP

The Series BCHP Calibration Test Pump 

is able to generate pressure

and vacuum for adjusting or calibrating pressure gauges, transmitters, or
switches.  The  pump  is  hand  operated  and  has  a  pneumatic  pressure
range of -28 in Hg to 870 psi (-0.95 bar to 60 bar). Dual pressure ports
allow  two  instruments  to  be  connected  to  the  pump  without  additional
fittings.  The  Series  BCHP  can  be  used  in  laboratories  and  production
areas.  The  fine  adjustment  valve  on  this  pump  ensures  precise
measurements.

PRODUCT OVERVIEW

SPECIFICATIONS:
Media:

Air.

Output Ranges:

-28 in Hg to 870 psi (-0.95 to 60 bar).

Process Connections:

1/4˝ female BSP

Gauge Connection:

1/2˝ female BSP.

Materials:

Anodized aluminum, brass, and ABS.

Weight:

8.4 lb (3.8 kg). 

Specifications - Installation and Operating Instructions

Series BCHP Low Pressure Hand Pump

11/32

[8.67]

9/64

[3.61]

11/32

[8.91]

3-3/16

[81.02]

43/64

[16.88]

APPROX. 18˝

1/4˝ FEMALE BSP W/ 

O-RING TEST GAGE

CONNECTION

29/64

[11.54]

7/8

[22.07]

61/64

[24.26]

29/64

[11.43]

1-13/32

[35.83]

3-3/16

[81.02]

1/2˝ FEMALE BSP

W/ O-RING

REFERENCE PORT

4-27/64

[112.40]

1-25/32

[45.16]

2-3/64

[51.85]

3-1/64

[76.57]

Figure 1

1. Pressure connector

for reference instrument
1/2˝ BSP female rotating

2. Fine adjustment knob
3. Pressure relief knob

4. Change-over switch for

pressure/vacuum generation

5. Pressure connector for test

instrument 1/4˝ BSP female
rotating

6. Pump handle

Do not use tape sealant to seal the ports, as it may damage the
pump.

NOTICE

MOUNTING INSTRUCTIONS

The reference instrument is to be mounted on the top connection of the
pump. Fastening with the knurled nut by hand is sufficient. An integral O-
ring  seals  the  reference  instrument  pressure  port  without  the  use  of
teflon tape.

The test instrument is attached to the connection at the end of the hose.
The  optional  NPT  adapter  set  includes  sealing  gaskets  and  different
adapters  to  best  fit  the  instrument  being  tested. The  tube  can  also  be
removed  to  attach  the  test  instrument  directly  to  the  pump  using  the
same adapter.

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176