Dwyer Instruments 490W Series Specifications-Installation And Operating Instructions Download Page 3

Select pair to 

measure pressure 

and flow

Touch to mark 

transducer as high

Touch SCAN 

button to scan for 

transducers 

Low side is 

automatically 

selected once 

high is chosen

Figure 3

Discovering the wireless pressure transducer

In the Sensor Discovery screen press the scan button to connect to transducers. 

Note: 

In this screen, up to three sets of transducers may be paired simultaneously 

to read differential pressure or up to six transducers may be may be paired to read 

single pressure.

Be  sure  the  selected  transducers’  MAC  addresses  have  the 

same maximum pressure. Each transducer is marked with a red 

(high) and blue (low) band to distinguish which transducer is high or low. Once the 

transducers are paired by MAC Address, select the High Side box for the transducer 

which has the red band. Its paired transducer will automatically be set to the low side. 

After  the  high  and  low  transducers  are  marked,  press  the  Selected  box  to  choose 

pairs to use.

NOTICE

The transducers will automatically go into low power/sleep mode 

to conserve battery energy. The transducers exit low power mode 

upon receiving communications from the Hydronic Application

NOTICE

Touch to 

start app

Figure 2

Touch to 

select valve 

make and 

model

Touch 

the NEXT 

button to 

proceed

Dwyer Hydronic App Operation

First turn on the handheld by pushing the power button (See Figure 1). Once fully 

powered on, unlock the handheld by swiping the screen to the right. The Dwyer 

Hydronic app is located at the bottom of the main screen. Click on the icon to open 

the app.

   

Valve Selection

Once the app is open, the Dwyer Hydronic App home screen will be shown. Press 

the START button. The Valve Selection screen will now be displayed. Press the make 

button to select a manufacturers valve. Then press the model button to select a valve 

model. In the list of valve manufacturers, there is an option to choose a manual valve. 

The flow coefficient (CV) is needed to use the manual valve, see Figure 2. Press the 

next button to proceed.

When the app is opened, permission messages will pop up. This 

can be bypassed by pressing ALLOW and can be permanently 

stopped by adding a Google Play

®

 account in the device’s settings menu.

NOTICE

Summary of Contents for 490W Series

Page 1: ...and power adapter for handheld test instrument Gorilla is a registered trademark of Corning Incorporated SPECIFICATIONS Wireless Distance Up to 65 19 8 m Service Compatible gases liquids Wetted Materials 316 SS PTFE brass Accuracy 2 of reading 1 psi Compensated Temperature Range 14 to 140 F 10 to 60 C Pressure Hysteresis 0 25 FS Pressure Range See chart Process Temperature Limits 4 to 185 F 20 to ...

Page 2: ...up to 2 years depending on usage When transducers are sent in for annual recertification the battery will be replaced Battery life depends on its capacity operating temperature and signal transmission interval When battery replacement becomes necessary use only a CR2050 or CR2032 type such as a Sony CR2050W or CR2032W NOTICE When the 490W is received the handheld test instrument will only have a p...

Page 3: ...cally go into low power sleep mode to conserve battery energy The transducers exit low power mode upon receiving communications from the Hydronic Application NOTICE Touch to start app Figure 2 Touch to select valve make and model Touch the NEXT button to proceed Dwyer Hydronic App Operation First turn on the handheld by pushing the power button See Figure 1 Once fully powered on unlock the handhel...

Page 4: ... pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with instructions may cause harmful interference to radio communications However there is no guarantee that interference will not o...