Dwyer Instruments 477AV-000 Specifications-Installation And Operating Instructions Download Page 3

Statistics

Minimum  and  maximum  readings  can  be  viewed  in  the  “STAT”  submenu.  While
they  are  being  viewed,  the  unit  of  measure  will  toggle  with  the  word  “MIN”  or
“MAX”. The minimum and maximum values can be reset by pressing and holding
the 

ENTER

button when viewing the values. They are also reset when the 477AV

is turned off, and when the mode of operation is changed. Here are the steps to get
to minimum and maximum values.
1. Press 

MENU

.

2. Scroll down to 

STAT

and press 

ENTER

.

3. Scroll to view the minimum and maximum values.

Calibration

If the measurements are not reading within specifications compared to a calibrated
reference,  the  zero  and  span  values  can  be  adjusted  in  the  calibration  menu.
When zero calibration is preformed, both ports should be vented to atmospheric
pressure.  When span calibration is performed, positive pressure equivalent to the
max end of the factory pressure range should be applied to the positive pressure
port with the negative port vented to atmospheric pressure.  This setting can be
cleared by a factory default.
1. Press  

MENU

and MEM will be displayed.

2. Press 

DOWN

arrow until CAL is displayed.

3. Press 

ENTER

.

4. Vent the pressure ports to atmospheric pressure.
5. Press and hold 

ENTER

.

6. Press 

DOWN

arrow.

7. Apply pressure to the positive pressure port.
8. Press and hold 

ENTER

.

Auto Power Off

The 477AV will turn off if no key is pressed for a set amount of time. The default
time is 20 minutes, but may be changed using the menu to any value from 1 – 20
minutes, or disabled. Here are the steps.
1. Press 

MENU

.

2. Scroll with the 

UP

and 

DOWN

arrows to “OFF” option and press 

ENTER

3. Using the 

UP

and 

DOWN

arrows, select a minute value for the auto-shut off.

4. Press 

ENTER

to lock in that value.

5. Press 

MENU

twice to return to the main screen. 

Sound

The 477AV can be set to beep during button presses. This feature can be turned
on or off using the “BEEP” menu option. Here are the steps.
1. Press 

MENU

.

2. Scroll down to the “BEEP” menu option and press 

ENTER

.

3. Using the 

UP

and 

DOWN

arrows, select “ON” or “OFF” and press 

ENTER

.

Factory Default

The  477AV  can  be  reset  to  factory  conditions  using  the  menu’s  "RSET"  option.
This will reset all settings, including a zero or span. Here are the steps.
1. Press 

MENU

.

2. Scroll using the 

UP

and 

DOWN

arrows to “RSET” and press 

ENTER

.

3. Scroll down to “YES” and press 

ENTER

.

Over Pressure/Under Pressure Alarm

If  over  pressure  or  under  pressure  is  applied  to  the  unit,  “OvFL”  or  “UnFL”  will
display respectively and an audible alarm will sound.  The threshold is ±10% of the
477AV's  range.  Do  not  exceed  the  maximum  rated  pressure  of  this  manometer;
doing  so  could  cause  a  permanent  damage  to  the  sensor,  rupture  the  housing,
and/or cause injury.

Battery Monitor

The 477AV monitors its battery voltage whenever it’s powered on.  If the voltage is
low,  the  “LOW  BAT”  icon  will  appear  on  the  display. A  weak  battery  can  cause
inaccurate  measurements.  It  is  recommended  to  replace  the  battery  as  soon  as
possible to ensure top functionality. Do not leave an exhausted battery in the unit
due to potential leakage.

Maintenance

The Series 477AV Handheld Digital Manometer is not field serviceable and should
be returned if repair is needed (field repair should not be attempted and may void
warranty). Be sure to include a brief description of the problem plus any relevant
application notes. Contact customer service to receive a return goods authorization
number before shipping. 

Advanced Menu

In  addition  to  the  features  accessible  directly  from  the  buttons,  there  is  an
advanced menu that allows access to stored values in memory, mode of operation,
digital  dampening,  measurement  statistics,  calibration,  power  saving,  audible
alarm, and factory defaulting the device.  The complete diagram of the advanced
menu can be found on the last page of this instruction manual or the back of the
unit.

1. Press and release the 

MENU

button to access the advanced menu.

2. Use the 

UP

and 

DOWN

arrow buttons to cycle through the menu options.

3. Press 

ENTER

once the desired option is selected.

4. Use the 

UP

and 

DOWN

arrow buttons to make changes to that option.

5. Press 

ENTER

to save the changes.

6. Press 

MENU

button to return to the menu or home screen.

Memory (MEM)

Saved readings can be viewed by pressing 

MENU

and then 

ENTER

. There will be

two options: “LOG” and “AVG”.

“LOG”  shows  saved  readings  in  the  order  they  were  stored.  “AVG”  displays  the
average value of the saved readings and is updated based on the content of the
log files. Empty memory locations are not shown. If no readings have been stored,
the  display  will  read  “----”.  When  the  memory  is  full,  the  display  will  flash  “ERR
FULL” and discard the reading. A reading cannot be stored when a menu is being
displayed.

Erasing Stored Readings

Stored readings can be erased using the menu.
1. Press 

MENU 

and MEM will be displayed.

2. Press 

ENTER

.

3. Press 

ENTER

as “LOG” is displayed on top of the screen.

4. Scroll using the 

UP

and 

DOWN

arrows to the value you wish to erase (or any 

if you want to erase all).

5. Press 

ENTER

.

6. This will take you to the “Erase” submenu. The “Erase” submenu has three 

options; “no” will do nothing, “OnE” will erase the most recently viewed reading, 
and “ALL” will erase all stored readings.

It  should  be  noted  that  the  measurement  mode  cannot  be  changed  unless  the
memory values are erased. (This prevents a user from accidently saving readings
with different units of measure.)

Changing Measurement Mode

Besides changing the measurement mode using the front buttons, it can also be
changed  in  the  menu.  The  477AV  can  be  set  to  display  pressure,  velocity,  or
volumetric flow. Once a mode is selected, the menu will continue to prompt for unit
of  measurement,  K-factor,  duct  shape  and  duct  size  when  applicable.  Duct  size
and shape are input in inches or centimeters, depending on whether the primary
unit  of  measure  is  metric  or  standard.  Settings  for  all  modes  are  retained  in  the
memory even if the mode is changed or the 477AV is turned off.

The mode 

cannot

be changed if there are readings saved in memory, or if the Hold

feature is on. (This prevents a user from accidentally saving readings with different
units of measure.) If a user tries to access the mode submenu when readings are
saved, the display will flash “ERR Log”. If a user tries to access the mode submenu
when the Hold feature is active, the display will flash “ERR Hold”.

1. Press 

MENU

and MEM will be displayed.

2. Press 

DOWN

arrow and MODE will be displayed.

3. Press 

ENTER

.

4. Using the 

UP

and 

DOWN

arrows, find the mode (velocity, pressure, or flow) you 

want.

5. Press 

ENTER

to confirm the desired mode.

6. The display will prompt for units of measurement. Scroll through the options 

with the 

UP

and 

DOWN

arrows until the unit of choice is found.

7. Press 

ENTER

.

Damping Function

The “DAMP” submenu compensates for the fluctuating readings by averaging input
values  to  display  a  more  stable  readout.  The  damping  value  can  be  set  to  any
number between 1 to 9. The default value is set to 4. Here are the steps to change
the damp value.
1. Press 

MENU

.

2. Scroll using the 

UP

and 

DOWN

arrows to select “DAMP” and press 

ENTER

.

3. Using the 

UP

and 

DOWN

arrows, choose a number in the range of 1 to 9 and 

press 

ENTER

.

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Summary of Contents for 477AV-000

Page 1: ...o 104 F 0 to 15 6 C and 25 6 to 40 0 C Pressure Hysteresis 0 1 of full scale Pressure Limits See chart Temperature Limits 0 to 140 F 17 8 to 60 C Compensated Temperature Limits 32 to 104 F 0 to 40 C Storage Temperature Limits 4 to 176 F 20 to 80 C Display 0 42 10 6 mm 4 digit LCD Resolution See chart Units of Pressure in w c ft w c in Hg psi oz in2 mm w c cm w c mm Hg mbar Pa kPa hPa Units of Velo...

Page 2: ...ult Display Backlight The 477AV has a display backlight to be used in the dark or in poor lighting conditions It can be turned on or off by pressing and holding the MENU and ENTER buttons at the same time The backlight setting is not saved after the 477AV is turned off Display Hold To hold or freeze a value press and hold down the HOLD button When the Hold function is enabled the pressure velocity...

Page 3: ...re is an advanced menu that allows access to stored values in memory mode of operation digital dampening measurement statistics calibration power saving audible alarm and factory defaulting the device The complete diagram of the advanced menu can be found on the last page of this instruction manual or the back of the unit 1 Press and release the MENU button to access the advanced menu 2 Use the UP...

Page 4: ... circ AREA rECt IN 1 00 IN 1 00 IN 1 00 BEEP OFF BEEP on MODE MODE PrES DAMP BEEP STAT 1 1 000 2 1 000 40 1 000 ERAS no ERAS OnE AREA dIA AREA ht AREA LEn KFAC 1 00 DAMP 4 MAX 1 000 MIN 1 000 OFF ZERO 1 000 MIN 20 SPAN 1 000 AVG 1 000 RSET RSET no RSET YES ERAS ALL LOG KFAC 1 00 CAL MPH unit FPM unit KN unit M H unit K H unit FPS unit M3S unit AREA OUAL IN 1 00 IN 1 00 AREA ht AREA LEn KFAC 1 00 P...

Reviews: