Dampening Function

The dampening feature allows the user to enter a dampening number from 1 to 16 (default

value = 2). Entering a larger number increases the amount of readings that are averaged for
each display update.

In order to access the dampening feature, follow the instructions below:

1. Press and hold the HOLD/MEMORY button. The upper right portion of the LCD scrolls
through a menu selection (HOLD, ST01, RD01, CLR, and DAMP). When “DAMP” is shown,
release the HOLD/MEMORY button. This selects the dampening feature.

2. Once “DAMP” is selected, a number is shown in the upper right portion of the LCD, along
with the current pressure reading. This number is the dampening number. Adjust the number
up by pressing the ZERO/STORE button or down by pressing the UNITS/LOC button. The LCD
update rate slows as the number increases from 1 to 16. Therefore, for best results, choose the
smallest number that provides a stable pressure reading.

Once the pressure reading is stable, press and release the HOLD/MEMORY button to store the
dampening value.

Overpressure Alarm

A visual indicator and audible alarm are provided to alert the operator that pressure has

exceeded the operating range of the unit. Exceeding the range will not damage it or affect
calibration as long as the maximum rated pressure is not exceeded.

Do not exceed the

maximum rated pressure of the manometer. Doing so will cause permanent damage
to the sensor, may rupture the housing and/or cause injury

.

The maximum pressure is

shown on the rear label and on page 1 of these instructions.

Low Battery Indicator

A weak battery can cause improper operation or inaccurate measurements. A low battery

indicator is provided on the display to show when the battery needs replacement. Although the
unit might appear to function and indicate properly, the accuracy of readings cannot be guar-
anteed when the LOW BAT indicator is illuminated. Replace the battery with a fresh one. Do not
leave an exhausted battery in the unit due to potential leakage.

MAINTENANCE

The Series 477A handheld digital manometers are not field repairable and should be returned

if repair is needed (field repair should not be attempted and may void warranty). Be sure to
include a brief description of the problem plus any relevant application notes. Contact customer
service to receive a return goods authorization number before shipping.  

Duracell

® 

is a registered trademark of The Gillette Company

Everready

®

is a registered trademark of The Everready Battery Company, Inc.

DWYER INSTRUMENTS, INC. 

Phone: 219/879-8000     www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.    

Fax: 219/872-9057          e-mail: [email protected]

D-26:D-26(478)  7/10/09  10:04 AM  Page 4

Summary of Contents for 477A-1

Page 1: ...ax 219 872 9057 e mail info dwyer inst com SPECIFICATIONS Service Air and compatible gases Wetted Parts Consult factory Accuracy 0 10 of full scale from 60 to 78 F 15 6 to 25 6 C 1 of full scale from 32 to 60 and 78 to 104 F 0 to 15 6 and 25 6 to 40 C Pressure Hysteresis 0 1 of full scale Pressure Limits See chart Temperature Limits 32 to 104 F 0 to 40 C Storage Temperature Limits 4 to 176 F 20 to...

Page 2: ...es a display backlight to allow use in the dark or in poor lighting con ditions Manometer must be switched off before this feature can be activated Next press and hold the ON OFF key down After about 1 second the backlight will come on and remain light ed for approximately 2 minutes after which it will turn itself off to conserve battery life Zeroing Pressure Reading Potential inaccuracy due to te...

Page 3: ...reading press and hold the HOLD MEMORY until ST01 is displayed then release the key Next press ZERO STORE key to save current reading to ST01 memory location A beep will sound indicating that the reading has been saved As each reading is saved the mem ory location display will advance to the next number To resume pressure measurement press the HOLD MEMORY key again Note that in the memory mode the...

Page 4: ... unit Exceeding the range will not damage it or affect calibration as long as the maximum rated pressure is not exceeded Do not exceed the maximum rated pressure of the manometer Doing so will cause permanent damage to the sensor may rupture the housing and or cause injury The maximum pressure is shown on the rear label and on page 1 of these instructions Low Battery Indicator A weak battery can c...

Reviews: