Dwyer Instruments 1430 Installation And Operating Instructions Download Page 4

When the readings are complete, the pres-
sure should be removed and the zero setting
of Microtector

®

rechecked.  Any change in the

zero position will indicate inaccurate readings.
Should this happen, the zero-set and pres-
sure measurement procedure should be
repeated.

Negative Pressure
or Vacuum Measurement

Zero the gage.  Connect the source of vacu-
um or negative pressure to the right-side
gage connection (Item 5) and proceed as
described under Positive Pressure
Measurement section.  Remember that the
pressure measured in this way is negative.

Differential Pressure  Measurement

Differential pressures may be measured by
connecting the higher (more positive) pres-
sure to the left connection (Item 2) and the
lower pressure to the right connection (Item
5).

Storage

Turn meter circuit switch to “off” position and
withdraw the point well clear of fluid (by turn-
ing micrometer clockwise) when gage is not in
use.  This will conserve the batteries and min-
imize build-up of oxides, etc., on the point.
Keep the unit covered and in an area free of
strong solvent fumes.

Maintenance

When the meter reading becomes reduced or
the pointer movement gets sluggish (with the
circuit on and the point in fluid), the following
should be done:

(1) Remove the point (by unscrewing) and
clean the tip lightly using fine crocus cloth.
Wipe off all grit and dirt with a clean rag;
reassemble and recheck meter operation.

(2) If the meter operation continues to be
sluggish, replace the size AA, 1.5 volt battery.
(Replace the battery at least once a year to
avoid deterioration of battery and damage to
gage.  Leakproof alkaline battery is recom-
mended.)

To replace the battery, remove center screw
(Item 10) located in the back of the electronic
enclosure.  Cover (Item 9) will come off,
exposing the battery.  Pull the old battery out
and push a new battery into the battery hold-
er with the positive (center) terminal to the
right (to the end marked with + on the holder).

If the fluid becomes contaminated and
requires replacement: empty old fluid from
gage; flush out with clear water and replace
with distilled water and Dwyer A126 fluores-
cein green color concentrate mixed with 3/4
oz. concentrate to each quart of water. 

CAUTION:

1. Do not substitute other gage fluids, as
proper gage operation depends on use of the
specified gage fluid to provide proper surface
tension, wetting ability and electrolyte capa-
bility with unity specific gravity.  

If the gage bore is very dirty, a mild soap solu-
tion may be used to aid in cleaning prior to
flushing with clear water.

2. Do not clean with liquid soaps, special sol-
vent, de-greasers, aromatic hydrocarbons,
etc.  Such cleaners and solvents may contain
chlorine, fluorine, acetone and related com-
pounds that will permanently damage the
gage and prevent proper operation.

©Copyright 2005 Dwyer Instruments, Inc         Printed in U.S.A. 10/05                       FR# 38-440190-00 Rev. 7

D-57  10/5/05  3:59 PM  Page 4

Summary of Contents for 1430

Page 1: ...red in accor dance with ASME B89 1 13 2001 and are traceable to a standard at the National Institute of Standards and Technology Three point mounting dual leveling adjustment and circular level vial assure rapid setup Durablock precision machined acrylic plastic gage body Sensitive 0 50 microamp D C meter acts as a detector and also indicates battery and probe condition Heavy 2 thick steel base pl...

Page 2: ...lifiers which act as a free run ning multivibrator operating at a frequency of approximately two kilohertz Completion of the A C circuit activates a bridge rectifier which provides the signal for indication on a sensitive 0 to 50 microamps D C microam meter 2 LOW PRESSURE FLUID LEVEL OFF ON On indication of contact the operator stops lowering the point and reads the micrometer which indicates one ...

Page 3: ...ing contact turn the point out of the fluid by turning the micrometer barrel counter clockwise to a reading of 010 or more Again watch the meter and this time lower the point by turning the micrometer barrel The point position where the meter pointer begins to move up scale is the zero position This position should correspond to the zero reading on the micrometer Adjust the point in relation to th...

Page 4: ...and dirt with a clean rag reassemble and recheck meter operation 2 If the meter operation continues to be sluggish replace the size AA 1 5 volt battery Replace the battery at least once a year to avoid deterioration of battery and damage to gage Leakproof alkaline battery is recom mended To replace the battery remove center screw Item 10 located in the back of the electronic enclosure Cover Item 9...

Reviews: