•   Flip the Level over 180° (see figure below) and 

press the 

CALIBRATE

 button. The number 2 will start 

to flash after a few seconds. This will indicate that 

the calibration has been completed.

•   The unit will automatically shutoff if it is stationary 

for three minutes.

NOTE: An incorrect calibration will occur if there is too 
much deviation in the horizontal calibration positions, or if 
the level has been shaken.

MAINTENANCE

Store and transport the digital level only in the supplied 
protective case. Keep the tool clean at all times. Do not 
immerse the tool into water or other fluids. Wipe off de-
bris using a moist and soft cloth. Do not use any cleaning 
agents or solvents. 

Checking the Measuring Accuracy

Check the accuracy of the digital level each time before 
using, after extreme temperature changes as well as after 
heavy jolts or impact.
The accuracy check should take place on a level and 
roughly horizontal surface. Power ON the digital level and 
place it on the horizontal surface. 
Select the unit of measure “°”. Wait for 10 seconds and 
note down the measured value.
Rotate the digital level by 180° around the vertical axis. 
Wait again for 10 seconds and note down the second 
measured value.
Calibrate the measuring tool only when the difference 
between both reading values is greater than 0.1°.

CALIBRATION

NOTE: When the ambient temperature changes consider-
ably, a calibration will be in order to maintain the accuracy 
of measurements.

•   Press the “

ON/OFF

” button to turn the unit on.

•   The LCD display will show the current angle.

•   Press and hold the “

ON/OFF

” button for 2 seconds to 

enter normal measuring mode.

•   In this mode, you need only to calibrate the horizon

-

tal surface.

NOTE: The 

 ° in/ft

RESET

 and 

HOLD

 buttons will not work 

in Calibration mode.

•   Press the 

CALIBRATE

 button and hold down for 4 

seconds. The LCD screen will display CAL1.

•   Put the digital level on a flat surface and press the 

CALIBRATE

 button again, the number 1 will start to 

flash. After a few seconds, the display will change to 
CAL2.

ABS

ABS