Usage
CAUTION
Risk of material damage due to at-
tachments becoming detached from
the compressed air pistol during
operation.
If the attachments become detached
from the compressed air pistol during
operation, this can cause material dam-
age.
i
Before using the unit, make sure that
the attachments are properly secured
on the compressed air pistol.
9.4 Notes on the compressed air
pistol
The compressed air pistol is available as
an accessory for SICOLAB mini ENDO
units and is operated with compressed
air.
Operating instructions for safe handling
of the compressed air pistol and cleaning
and reprocessing instructions are availa-
ble as a download
"Operating instruc-
tions for cleaning pistols made of
stainless steel"
from
www.reinigung-
spistolen.de
.
NOTICE
Excessively high pressure on the
unit can cause damage to the endo-
scopes.
If the pressure settings on the unit are
higher than the pressure specifications
provided by the endoscope manufactur-
er, the endoscopes can be damaged.
i
Adjust the pressure settings in ac-
cordance with the information provid-
ed by the endoscope manufacturer.
Due to design-related differences in the
channel diameters of the endoscopes to
be dried, the drying times can vary.
Note: When drying medical devices such
as endoscopes, always follow the appli-
cable specific regulations from the manu-
facturer or from the Robert Koch Insti-
tute.
WARNING
Risk of injury due to improper use
of the compressed air pistol if the
compressed air pistol is pointed at
parts of the body or orifices while it
is used.
The pressure of the air flow can cause
injuries.
i
During operation, the compressed air
pistol must not be directed at parts of
the body or orifices.
EN