Adjust the time and tap
Save
. Tap
to go
back.
Automatic adjustment
In order to use automatic time setting,
DHCP
must be activated (see "9.7 Con-
❯
Tap
Time Zone
.
❯
Adjust the time zone and tap
Save
.
9.4
Adjusting the temperature
❯
Tap
Sealing Parameters > Temperature
.
❯
Adjust the temperature.
❯
Tap
Save
.
Recommended sealing temperatures:
– 180 °C is recommended for Hygofol transpar-
ent sterilisation packaging.
– For sterilisation packaging from other manu-
facturers, the recommended sealing tempera-
ture should be checked with the relevant man-
ufacturer.
After adjusting the temperature, perform a
test sealing process.
9.5
Set the speed for the belt
transport
The speed of the belt transport can be adjusted
between
Slow
,
Medium
and
Fast
.
❯
Tap
Sealing Parameters > Speed
.
❯
Select the required speed, tap e.g.
Medium
.
❯
Tap
to go back.
❯
Perform a sealing test, see "14.2 Daily function
assessment").
After changing the speed it may be nec-
essary to adjust the sealing temperature
(see "9.4 Adjusting the temperature").
9.6
Set standby time
❯
Tap
System Settings > Standby
.
❯
Tap
+
or
-
and adjust the time.
It is set to 30 minutes by default.
❯
Tap
Save
.
9.7
Configuring the network
After the network settings have been
changed it is necessary to restart the
device. This saves the changed settings.
Requirements:
ü
Access level
Administrator
or
Service Techni-
cian
selected.
If DHCP is activated then the settings are
applied automatically.
Deactivate DHCP if you wish to adjust the
settings manually.
❯
Tap
System Settings > Network
.
❯
Deactivate
DHCP
.
❯
Tap
IP Address
, enter the data and tap
Save
.
❯
Tap
Net Mask
, enter the data and tap
Save
.
❯
Tap
Gateway
, enter the data and tap
Save
.
Syslog settings
If the unit is monitored using third-party docu-
mentation software for documentation of the
sealing processes, Syslog settings will still need
to be adjusted for it.
Syslog IP address:
IP of the computer on which
the documentation software is running.
Syslog port:
The port via which the units are
communicating (default setting: 514).
It is important that no other program is allowed to
access the same port on the computer on which
the documentation software is installed, as other-
wise that port will be blocked and the documen-
tation software will not be able to receive any
data.
❯
Tap
System Settings > Network > Syslog IP
Address
, enter the data and tap
Save
.
❯
Tap
Syslog Port
, enter the data and tap
Save
.
Syslog Heartbeat is activated at the fac-
tory, which allows the network stability to
be improved.
Assembly
6024100132L02 1906V003
15
EN