Plug a network cable into the network connec-
tion on the unit (optional when using
e.g. Tyscor Pulse) and into the network socket.
8
Commissioning
❯
Insert the SD card into the unit.
❯
Switching on the unit:
Press the On/Off switch
(blue LED lights
up).
❯
A signal sounds, the blue LED flashes and the
device is heated to the pre-set temperature.
Once the selected temperature is reached, the
blue LED goes out and the device is ready for
operation.
❯
Set up the basic settings for the device – e.g.
temperature, date and time. The firmware ver-
sion can be checked under
Device Information
> Device Data
The temperature of the device is pre-set
to 180°C for Hygofol transparent sterilisa-
tion packaging from Dürr Dental.
The speed of the belt transport is pre-set
to
Medium
.
❯
Carry out a function check of the device.
❯
Perform a sealing test.
If the test seal is not OK, adjust the tempera-
ture and/or speed as required.
8.1
Monitoring the unit via the
network
If the device is connected to the network then it
can be monitored with the VistaSoft Monitor soft-
ware.
VistaSoft Monitor shows the current status along
with messages and faults on the device.
In addition, VistaSoft Monitor also helps with
maintenance of the device – the software warns
when maintenance work is due.
For further information please refer to the
VistaSoft Monitor manual (order number
2110300001).
Assembly
6024100132L02 1906V003
13
EN