![Durr Dental 6024100001 Installation And Operating Instructions Manual Download Page 16](http://html.mh-extra.com/html/durr-dental/6024100001/6024100001_installation-and-operating-instructions-manual_2542885016.webp)
8.2
Handover of the device
A factory calibration is performed on the
device at the factory. A printout of these
test results is included in the supplied
documents. This factory calibration does
not replace the initial validation of the
sealing process after commissioning at
the place of installation. To do this, a per-
formance qualification of the sealing pro-
cess must be performed after commis-
sioning (see "14.3 Annual Performance
Qualification").
Correct handover and installation of the device
forms part of the validation process during the
Installation Qualification (IQ). This process must
be documented like every part of the validation.
Please use the enclosed handover record or
download and print it out from www.duerrden-
tal.com and then fill it in.
A new
Installation Qualification
always
needs to be carried out if major changes
are made to general conditions.
9
Configuring the unit
9.1
Selecting the access level
The adjustment options depend on the selected
access level.
Only a qualified expert is permitted to
change the settings in the Service Techni-
cian access level.
Access level
Rights
Operator
– Querying device information
– Sealing parameters
Administrator
In addition to the User rights:
– System settings (administra-
tive)
For operators, e.g. hygiene
officer
Service Tech-
nician
In addition to the Administrator
rights:
– System settings (all)
For qualified Service Techni-
cians
❯
Tap
System Settings > Access Level
.
❯
Select the access level.
❯
Tap
to go back.
9.2
Language selection
❯
Tap
System Settings > Language
.
Language selection:
❯
Select the language and tap
Save
. Tap
to go back.
9.3
Set the date and time.
The date and time must be set correctly on the
device in order for the batch documentation to
be correct.
You can manually set the date and time or, if the
device is connected to the Internet, you can
obtain the correct time and date from a time
server.
Manual adjustment
❯
Tap
System Settings > Date / Time
.
Manual adjustment of date and time:
❯
Tap
Date
.
❯
Adjust the date and tap
Save
. Tap
to go
back.
❯
Tap
Time
.
Assembly
14
6024100132L02 1906V003
EN
Summary of Contents for 6024100001
Page 1: ...EN Hygopac View Installation and operating instructions 6024100132L02 6024100132L02 1906V003 ...
Page 2: ......
Page 34: ......