22
23
EMC Guidance
Table 1 – Guidance and MANUFACTURER’S declaration –
ELECTROMAGNETIC EMISSIONS – for all ME EQUIPMENT and ME
SYSTEMS
Table 2 – Guidance and MANUFACTURER’S declaration – electromagnetic
IMMUNITY – for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic emissions
The BS901 is intended for use in the electromagnetic environment specified below. The customer or
the user of the BS901 should assure that it is used in such an environment.
Emission test
Compliance
Group 1
Class B
Not Applicable
Not Applicable
RF Emission
CISPR 11
RF Emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Electromagnetic environment – guidance
The BS901 uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The BS901 is intended for use in the electromagnetic environment specified
below. The customer or the user of the BS901 should assure that it is used in
such an environment .
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
±2 kV for
power supply lines Not applicable
Not applicable
Not applicable
±1 kV line(s)
to line(s)
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3A/m
NOTE U T is the a.c. mains voltage prior to application of the test level.
IEC 60601
test level
Compliance
level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3A/m
Power frequency
(50Hz) magnetic
field IEC
61000-4-8
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines IEC
61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5