VistaCool is a patent-pending system designed to eliminate the
need for autoclave condenser/waste bottles and condensation
tanks by cooling the exhaust water and automatically sending it
directly down the drain� With the use of a simple adapter kit, the
VistaCool Model V7502 is designed to enable the Midmark/Ritter
M9 and M11 autoclaves to use fresh deionized/distilled water for
each sterilizing cycle—converting these chambered autoclaves to
operate like quick-turnaround, cassette systems�
Introduction
VistaCool can be used with:
Midmark/Ritter M3
Midmark/Ritter M9*
Midmark/Ritter M11*
SciCan STAT
IM
2000
SciCan STAT
IM
5000
We recommend that a professional technician or plumber familiar with dental/medical offices perform the
installation since interface with a cold water line and drain is involved� He/she should be familiar with local
plumbing codes and techniques for successful dental/medical equipment installations� Failure to install
correctly could result in water leaks, leading to possible water damage� It is the responsibility of the system
owner to validate the use of all equipment within its installed environment�
Please keep the Installation Guide & Owner’s Manual handy for future reference and ensure that anyone
responsible for operation and maintenance of the system is familiar with all details contained in this manual�
Please return the Warranty Registration form or register online immediately upon installation�
IMPORTANT INFORMATION
For System Owners
Please read this entire manual before proceeding with installation and operation, and always follow local
plumbing codes�
VistaCool systems have only been verified for installation on the compatible autoclaves (listed above)�
Please check with the manufacturer of other brands of autoclaves or contact Vista Research Group before
connecting to other than SciCan or Midmark/Ritter autoclaves� See page 13 for autoclave-specific instructions�
VistaCool Model V7502 has been verified for installation on Midmark/Ritter M9 & M11 units with the
use of the S7560 adapter kit� Only M9, M9D, M11 and M11D (-02X, -03X) units with
green
displays can
be connected to VistaCool� First generation models with
red
displays (-0XX) are not compatible�
The VistaCool tank must be installed so that the top of the water tank is ABOVE the connection to the
drain adapter� Please see page 12 for important tank positioning details�
Check for leaks and run the autoclave(s) for at least two consecutive cycles after installation to
ensure the system is operating properly�
IMPORTANT INFORMATION
For System Installers
* Requires Model V7502 and S7560 adapter kit