5 | Installation
D-R 290
99
DANGER
High voltage. Risk of fatal injury due to electric shock!
Touching live parts poses an immediate risk of fatal injury.
▶
Permit only qualified electricians to work on electrical equipment.
▶
Before opening the casing of a device or removing a guard protecting against
touching it, deenergise the device, test it to ensure it is electrically dead and se
cure it against switching on again.
In all installations, ensure compliance with the applicable local regulations and any
supplementary regulations issued by the local power supply companies.
1.
Loosen the four (captive) screws and open the supply unit.
Spring terminals are provided in the supply unit for connecting the data cable
screens. These terminals can be released using a hexagon socket key, moved to
the available tapped holes on their rail (
– 1) if required and also changed
to a different direction.
1
Tapped hole
2 Recess in the spring terminal
Fig. 5.33: Connecting the cable screen
1.
Strip back approx. 20 mm of the screen.
2.
Using a suitable tool (e.g. a screwdriver), depress the spring terminal into the re
cess (
– 2) provided for this and insert the screen between the jaws of the
terminal.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...