5 | Installation
98
D-R 290
The supply unit is installed with the cable glands facing downwards. Unused openings
must be sealed to prevent the ingress of moisture and protect against corrosion inside
the housing.
Fig. 5.31: Supply unit cover assembly
Carefully note the alignment of the casing cover (
symmetrical in orientation. If the alignment is wrong the gasket will be damaged and
the supply unit will no longer satisfy protection type IP 66 (NEMA 4.4x).
Fig. 5.32: Attaching the retaining plates
If desired, the retaining plates can also be attached to the housing in the vertical align
ment. They then project half way left and right beyond the housing (for dimensions see
Electrical connection to the D-TB 100 supply unit
The supply unit sends the measurement data prepared by the measuring head to the
customer interface and serves as power supply for the connected sensor.
When connecting the supply unit, refer also to section
5.2.2 Instructions for planning
the electrical connections to the system [
5.7.2
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...