5 | Installation
68
D-R 290
Within the measurement duct…
▶
…the distribution of the dust and smoke within the gas should be as homogeneous
as possible
▶
…the dewpoint of the flue gases should not be undershot
▶
...the path of the light beam through the gas to be measured should be as nearly
horizontal as possible.
The measuring point should
not be just before or just after bends in the duct or
changes in its cross-section. The upstream section (duct section leading up to the
measuring point) should be at least 5 × D (D = internal diameter of the duct) and the
downstream section (duct section leading away from the measuring point)) should be
a). If no such location is available, the upstream section (A)
should be longer than the downstream section (see
On flues with round internal cross-sections, "D" means the internal diameter. In the
case of square or rectangular cross-sections, the hydraulic diameter is used:
Equation
10
F
Area
U
Scope
Table 5.2: Equation 10
The installation location should be as free from vibrations and oscillations as possible.
If this is not possible, it is recommended that the installation it keep away from
draughts that might induce vibration, so that the measuring system is not subject to
stresses.
Fig. 5.2: Measuring point
If a measuring point is sought between a filter and an induced draught, then the instal
lation point should be placed more towards the induced draught and less towards the
filter.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...